10 Error and system messages
Error messages during measurement
Error messages during calibration
Display
Cause
Remedy
Battery voltage low,
the function can still be guaranteed only
for a short time.
Insert new battery
Battery is used up
Insert new battery
During mains operation: incorrect voltage
Test power supply unit, replace if necessary
No display or
muddled charac-
ters, device does
not respond to
keys being
pressed
Battery voltage too low
Insert new battery
During mains operation: power supply
unit defective or incorrect voltage/polarity
Test power supply unit, replace if necessary
System error
Remove battery or power supply unit, wait
briefly before reconnecting
Device error/device defect
Send in device for repair
Measuring range exceeded
Check if the measured value is above the admissi-
ble measuring range of the combination electrode
Combination electrode is defective
Send in device for repair
Measuring range not reached
Check if the measured value is above the admissi-
ble measuring range of the combination electrode
Combination electrode is defective
Send in device for repair
System error
Send in device for repair
Measuring range exceeded or not
reached by a significant margin
Check whether the measured value is within
the admissible measuring range of the combi-
nation electrode
Preset calibration interval elapsed
Calibrate device
Last calibration invalid
Display
Cause
Remedy
The neutral buffer is not admissible:
•
Combination electrode is defective
Clean combination electrode, recalibrate.
If error reoccurs: replace combination elec-
trode
•
Wrong buffer solution
Always use neutral buffer as first solution
(exception: 1 point calibration).
•
Buffer solution used up
Use new buffer solution
The slope is too low:
•
Combination electrode is defective
•
Buffer solution used up
Replace combination electrode
Use new buffer solution
The slope is too high:
•
Combination electrode is defective
•
Buffer solution used up
Replace combination electrode
Use new buffer solution
Wrong
calibration
temperature
Only carry out calibration between
0 and 60 °C
All manuals and user guides at all-guides.com
Summary of Contents for 202710
Page 2: ...All manuals and user guides at all guides com ...
Page 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Page 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Page 57: ...All manuals and user guides at all guides com ...