41
8 Device settings
8.4
Parameterizing the device
8.4.1
Submenu "SEt Corr": offset/zero point correction and slope cor-
rection of the voltage and temperature measurement
Overview of the parameters
After calling up the setting menu with
and selecting the submenu
"SEt Corr", use
to select the parameter to be configured.
Parameter "OFFS": offset/zero point correction of the voltage measurement
✱
Use the
and
keys to set the zero point correction:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameters
(3)
"OFFS mV": offset/zero point correction of the voltage measurement, See
page 41
(4)
"SCAL mV": slope correction of the voltage measurement, See page 42
(3)
"OFFS °C": offset/zero point correction of the temperature measurement, See
page 42
(4)
"SCAL °C": slope correction of the temperature measurement, See page 43
CAL
3
(1)
(2)
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
mV
mV
(4)
(3)
Menu/parameter
Setting option
Meaning
oFF
no zero point correction of the voltage
measurement
-
10.0 to +10.0 m
V
zero point correction of the
voltage measurement in mV
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
all-guides.com
Summary of Contents for 202710
Page 2: ...All manuals and user guides at all guides com ...
Page 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Page 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Page 57: ...All manuals and user guides at all guides com ...