71
S 20.00
ID 0
Caution:
•
Choose the detergent according to the bath liquid which is used.
E.g. Ethanol, Silicon remover can be used.
•
Take care the room where cleaning takes place is well ventilated.
•
Set the working temperature setpoint.
Recommendation: 20 °C
Cleaning the inside:
If you change the bath liquid or want to use a different
one, carefully clean the parts in contact with the bath
liquid. The heat exchanger, internal reservoir,
connecting tubing and external system should be
rinsed several times using cleaning liquid.
Preparations:
1. Drain the temperature system.
See chapter 5.9. Draining
2. If only the Presto temperature system is to be
cleaned, connect the pump connectors (13, 14)
using a tubing.
3. Connect a tubing to the connector (16) "Reservoir
extern". First hold the end of the tubing towards
position 16.1, to prevent the cleaning liquid from
flowing down instantly.
4. Hold the tubing from the overflow connector (15)
into a suitable vessel.