Menu functions
50
8.4. Integrated
programmer
The integrated programmer allows any desired temperature
program sequences to be realized. Such a temperature sequence
is called profile. A profile consists of individual sections defined
by duration (t:) and target temperature. Target temperature is the
setpoint (T:), that is achieved at the end of a section. The
programmer uses time and temperature difference values within a
section to calculate a temperature ramp.
•
Press
enter
to select the submenu "Int. Programmer".
•
Use the up/down cursor keys
to select the desired
option. Then press enter
to open.
A flashing segment indicates that a number or value needs to
be entered.
Edit Compile
profiles
Display
sections
Delete
Delete sections
Set clock
Set the real time on the instrument
Examples:
Setting clock
The integrated clock allows starting a profile at any date and time.
The clock is preset in the factory.
•
Lines 1 to 3:
Check for correctness of the preset date and time and correct
if necessary.
The time is diplayed permanently in line 4.
•
Use the numerals to set time, date and year and set each
entry with enter
.
•
Press
escape
to return to the previous menu level.