11.3.3 Function codes
Name
Code
(dec.)
Code
(hex.)
Explanation
Read Holding
Registers
03
03
Read several holding registers
Write Single
Registers
06
06
Write a single holding register
Write Multiple
Registers
16
10
Write several consecutive holding registers. This
function can be used to set values that use more
than one register.
Read/Write
Multiple
Registers
23
17
Combination of read and write operation
Read Input
Register
04
04
11.3.4 Holding registers
Register
address
Protocol
address
Data type
Explanation
Adjustable range
40001
0
ushort
Start/stop the device
0: Device is in standby
1: Device is started
40002
1
ushort
Type of actuating variable
0: Controller
1: Digital
2: EPROG
40003-
40004
2-3
float
Setpoint temperature
min. - max.
40005
4
short
Actuating variable via
Modbus
-100 … 100
40011
10
ushort
Internal/external
temperature control
0: Temperature control
in the internal bath
1: Temperature control
with external Pt100