Operation
69
The programmer is activated. It starts either immediately or at the
programmed start time on the specified start date and goes through the
specified number of runs one after the other. After the last run, the device
switches to the defined final state.
When the programmer is active, the
[Quick access programmer]
button
is displayed at the bottom of the home screen. Use this button to stop or
end the programmer, or edit the active profile.
8.10 Configure signal outputs
8.10.1 Set up Stakei outlet
A solenoid valve connected to the Stakei output can be assigned various
functions depending on the application.
The device is switched on.
A solenoid valve is connected to the Stakei output.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Connect unit>
then menu option
<Interfaces>
.
3. Use the
[Control output]
button to select the desired function for the
connected solenoid valve.
The Stakei output is set up and active.
8.10.2 Configure signal outputs of the REG/EPROG plug
The REG/EPROG socket provides three channels as signal outputs. The voltage
or current signals emitted via these channels can be externally processed.
The following values can be output:
Setpoint: Specified setpoint
Internal: Internal temperature sensor value
External: External Pt100 temperature sensor value
Power output: Value of the variable
The device is switched on.
The electronics module with analog connections is mounted.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Connect unit>
.
3. From the submenu
<Interfaces>
, select the menu option
<Analog
module>
.
4. For
[Channel 1]
select the output value setpoint, internal, external or
capacity.
5. Enter the lowest output value as the 0 V value.
6. Enter the highest output value as the 10 V value.
7. Repeat the steps 4 to 6 for
[Channel 2]
.