Operation
72
8.12.2 Adjust the internal temperature sensor - multi-point calibration
This section describes how to adjust the internal temperature sensor of the
device with multiple correction points. The adjustment can be performed with
up to ten correction points.
The device is switched on.
The bath tank is at least two thirds full.
A calibrated reference thermometer is available.
1. Initially proceed in the same way as for the 1-point adjustment.
2. After confirming the first correction value, go to the start screen and enter
the setpoint temperature for the second correction point.
3. Wait until the temperature in the bath has stabilized and is shown on the
display of the circulator.
4. Switch to the submenu
<ATC internal>
and press on the
[Add correction point]
button.
5. Tap the added
[Correction point 2]
button.
6. Enter the displayed reference value for the second correction point in the
input window and confirm your input.
The 2-point adjustment is completed. The correction values are applied
when starting the temperature control application.
7. For another correction point, set the next desired setpoint temperature and
wait until the temperature in the bath has stabilized.
8. Repeat steps 4 to 6 for the next correction point and then confirm your
input.
Up to ten correction points can be created for the adjustment.
The multi-point adjustment is completed. The correction values are applied
when starting the temperature control application.
The internal temperature sensor is adjusted.