Error messages
36
Error messages
The following table contains some of the possible
error messages with the associated warning messages
and resolutions.
All warning messages are indicated on the display and
accompanied by an audible signal. To mute the audible
signal, press
once. The error message will still be
visible.
Error messages
When the following error messages appear, the unit
(heater and circulation pump) will shut off. You must
remove the cause of the problem before you can
switch the unit back on.
Shown on the display
What should I do?
E 01: LOW WATER LEVEL
ALARM
Not enough water in the bath tank.
Switch off the unit at the power switch, add
water, and switch it back on.
If there is enough water in the bath tank:
The float switch is defective (transport dam-
age, etc.). Sent the unit to an authorized
fu-
sionchef
by Julabo
repair facility.
E 05: WORKING SENSOR
ALARM
Please contact our service department.
E 06: SENSOR
DIFFERENCE ALARM
Please contact our service department.
E 07: INTERNAL HARD-
WARE ERROR
Please contact our service department.