Changing the water
31
Care and maintenance
The circulator is designed for continuous operation
under normal conditions. Regular maintenance is not
required. However, you should perform the following
cleaning tasks each day.
Changing the water
The circulator is suitable only for use with sof-
tened/decalcified water. The water in the stainless
steel bath should be changed regularly. Our recom-
mendation:
Change the water daily
for the following reasons:
•
Frequent refilling increases the concentration of
minerals in the water, which can deposit inside
the bath and on immersed parts of the circulator
in the form of a crust.
•
Vacuum packaging may be slightly dirty; these
particles will be released into the warm water
bath.
Draining combination units
To drain a
Pearl
Edition XS, S
,
M
,
L
, or
XL
com-
bination unit, use the integrated drain tap.
1. Switch off the unit and pull the power plug.
2. Place the unit over a sink or a container large
enough to hold all of the water.
3. Open the drain tap and allow the water to flow
out.