ix
side facing the chest. If the pads are not securely attached to the
chest, the AED may not function. Paste the pads exactly at the
positions indicated on the pads, If the chest is wet with water, wipe
dry with a dry towel and the like, and then paste the pads. If there is a
pacemaker or implantable cardioverter defibrillator (ICD), paste the
pads at least 3cm away from them. If a medical patch or plaster is
present, peel it off and then paste the pads. If the injured or ill person's chest hair is thick, paste
the pads on the chest hair once, peel them off to remove the chest hair, and then paste new
pads.
c) Some AED models require to connect a connector by following voice prompts.
d) Do not put child pads on adults (older than 8 years).
12
.
Electrocardiogram analysis
a) The AED automatically analyzes electrocardiograms. Follow the
voice prompts of the AED and ensure that nobody is touching the
injured or ill person while you are operating the AED.
b) On some AED models, you may need to push a button to analyze
the heart rhythm.
13
.
Electric shock (defibrillation)
a) If the AED determines that electric shock is needed, the voice
prompt saying, "Shock is needed" is issued and charging starts
automatically.
b) When charging is completed, the voice prompt saying, "Push the
shock button" is issued and the shock button flashes.
c) The first-aider must get away from the injured or ill person, make
sure that no one is touching him/her, and then press the shock button.
d) When electric shock is delivered, the body of the injured or ill person may jerk.
14
.
Resume cardiopulmonary resuscitation (CPR).
Resume CPR consisting of 30 chest compressions
and 2 rescue breaths by following the voice prompts
of the AED.
15
.
Automatic electrocardiogram analysis
a) When 2 minutes have elapsed since you resumed cardiopulmonary resuscitation (CPR), the
AED automatically analyzes the electrocardiogram.
b) If you suspended CPR by following voice prompts and AED voice prompt informs you that
shock is needed, give electric shock again by following the voice prompts.
If AED voice prompt informs you that no shock is needed, immediately resume CPR.
Press the shock button.
Summary of Contents for JAN-7201
Page 2: ......
Page 24: ...xxii CWA 246 26inch Display Unit Mount Kit Warning Label ...
Page 25: ...xxiii CWA 245 19inch Display Unit Mount Kit Warning Label ...
Page 28: ...xxvi Warning Label NKE 1632 Radar Antenna Warning Label NKE 2632 2632 H Radar Antenna ...
Page 30: ...xxviii NTG 3230 Transmitter Receiver 30 kW ...
Page 33: ...xxxi CWA 245 Display Unit Mount Kit CWA 246 Display Unit Mount Kit ...
Page 76: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 13 ...
Page 78: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 15 ...
Page 80: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 17 ...
Page 82: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 19 ...
Page 84: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 21 ...
Page 86: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 23 ...
Page 88: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 25 ...
Page 90: ...2 Installation of Scanner Unit 2 2 Installation for the specified scanner model 2 27 ...
Page 92: ...2 Installation of Scanner Unit 2 3 Installation of Transmitter Receiver 2 29 ...
Page 94: ...2 Installation of Scanner Unit 2 3 Installation of Transmitter Receiver 2 31 ...