Choose Unit of Standard Samples
Move the cursor onto “
Unit”
by pressing
[»]
and
[«]
, followed by
[ENTER]
, 8 units
are under your selection (Fig 13).
Move the cursor onto the unit you need, then press
[ENTER]
. You can also press the
number ahead of the unit directly by means of the numeric keypad.
If you don’t want to select any item, press
[RETURN]
to go back to the last screen.
Set Quantity of Standard Samples
Move the cursor onto “
Number”
by pressing
[»]
and
[«]
, followed by
[ENTER]
.
(Fig. 14)
Input the number of your standard samples by means of the numeric keypad. After
pressing
[ENTER]
, it returns to the setting screen.
Note:
Number range is from 1-9, other number is invalid.
Set concentration of standards
Before this step, you must blank first. Pull the blank cuvette into the light path in the
pre-testing interface, and press
[ZERO]
.
Move the cursor to
Concentration
, and press
[ENTER]
. (Fig. 15)
Pull sample 1 in the light path, input the concentration of sample 1 by means of
the numeric keypad, then press
[ENTER]
. It goes into the next setting interface of
Figure 13
Figure 14