
MANUAL DE INSTRUCCIONES CODIGO 80243 REV A Enero /11
(Sujetas a modificaciones sin previo aviso) Pag.: 38
J
.P. SELECTA s.a.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 93 770 08 77 Fax 34 93 770 23 62
e-mail: [email protected] - website: http://www.jpselecta.es
Enter Quantitative Mode
Move the cursor onto the
Quantitative Mode
, and press
[ENTER]
to go into the
method choosing interface (Fig. 9).
2-1. Standard Curve Method
It’s a method to establish a standard curve by measuring a group of standard samples.
Move the cursor onto the
Standard Curve
, then press
[ENTER]
, and it will go into
the pre-testing interface (Fig.10).
Set Wavelength
Press
[GOTO]
in the pre-testing interface, input the testing wavelength by pressing
the numeric keypad (Fig.11).
Blank
Pull the blank solution in the light path, and then press
[ZERO]
.
Set Parameters
Press
[SET]
in the pre-testing interface. It goes into parameters setting interface
(Fig.12). You should set
Unit
, the
Number
of standard samples, and
Concentration
of standard samples before testing.
Figure 9
Figure 11