Enter Quantitative Mode
Move the cursor onto the
Quantitative Mode
, and press
[ENTER]
to go into the
method choosing interface (Fig. 9).
2-1. Standard Curve Method
It’s a method to establish a standard curve by measuring a group of standard samples.
Move the cursor onto the
Standard Curve
, then press
[ENTER]
, and it will go into
the pre-testing interface (Fig.10).
Set Wavelength
Press
[GOTO]
in the pre-testing interface, input the testing wavelength by pressing
the numeric keypad (Fig.11).
Blank
Pull the blank solution in the light path, and then press
[ZERO]
.
Set Parameters
Press
[SET]
in the pre-testing interface. It goes into parameters setting interface
(Fig.12). You should set
Unit
, the
Number
of standard samples, and
Concentration
of standard samples before testing.
Figure 9
Figure 11