Unit Operation
Unit Operation
Unit Operation
19
I
n the press key insert battery into the test mode at the same time,
according to an inferior version available algorithm and software
version number, alternating press
and V may continue to see test.
Λ
Power Off
-Transfer your readings
1. As soon as your measurement is complete, open the app on your smart phone to transfer
your reedings.
Notr: On the paired smartphone, Bluetooth must be enabled.
2. You can view your blood pressure readings on the app.
The MAC address
20
Unit Operation
Note: Only Service personnel permitted to access to this mode, the mode unavailable in
normal use.
Bluetooth requirements
The monitor requires a device with:
. Bluetooth 4.0 or later
. Android 5.0 or later
. IOS 9.0 or later
And works with:
. iphone , iPod, iPad
. Android Phones and Tablets
-Pairing
your monitor with a
S
mart Device
Wireless communication
Frequency range : 2.4 Ghz (2400-2483.5 MHz)
Modulation : GFSK
Antenna gain:0.5dBi
1. Download the free “JoyHealth”
App: On your mobile phone or table
go to www.sejoy.com.
2.Open the App on your phone or tablet. If requested, you should enable Bluetooth on your
device. You can enable Bluetooth under the Settings menu on your smart phone or table.
3. Create a new user login, or login with your existing user name and password.
4. Selection device “Blood pressure monitor”.
Bluetooth connection
-Using
for
the
first
time
1. Long press the " " button on your monitor until the word "bE" flashes on your monitor .
2. Open the APP and click "pair device" button, a popup window will appear. Click "Search device"
button to display all sphygmomanometer devices(The display information includes the device model
and MAC address), then you can find the sphygmomanometer through the "device Name", click the
pictureA film or text can be used to pair the device.
3. After the pairing is complete, the device interface switches from boo interface, indicating that
the pairing is successful.
Troubleshooting
Abnormal
phenomenon
Processing method
Abnormal
sphygmomanometer
Cause analysis
Note: If you cannot solve the abnormal situation by yourself, you can consult the manufacturer or
the manufacturer's designated unit by phone. It is forbidden to disassemble and repair without
permission. If necessary, professional maintenance personnel can ask the manufacturer for the list
of components and circuit schematic diagram.
The following table shows the error signs that may occur during measurement, possible causes
and handling methods. Please measure again using the correct method
Er1
Er2
Er3
Error display
The cause of the problem
The solution
Can't detect high and low
pressure
Cuff too loose or loose
Improper compression caused
by arm or body movement
The pressure exceeds 300mmHg
Please fasten the cuff before
measuring
Hold the arm or body still and
measure again
Battery dead
Replace the battery or connect the
power adapter (if any).
Er4
The pressure exceeds 15mmHg
for 3 minutes
Check whether the cuff is knotted or the
vent valve is blocked. If the problem
persists, contact the manufacturer
Move the arm during measurement or
Electronic sphygmomanometer
Speaking, nervous or emotional
during measurement
There is interference in charging process
or improper operation in measuring
process
Incorrect measurement posture
Stay quiet, keep your arm steady, and
do not move the monitor
Instead of talking, take deep
breaths to calm your mood and
relax your body
Adjust posture, see
"Blood pressure gauge Wearing”
See operation Instructions.
The armband is tied too tight or too loose,
Or the arm strap is tied incorrectly;
Roll the armband correctly
Please fasten the cuff before
measuring
Please fasten the cuff before
measuring
Note: Memory cannot be recovered once it has been deleted.
If you want to delete the other set of memories, please user memory group set up in front of the boot,
specific operation as shown in the "system Settings" in " User memory group Settings ";
Static Pressure Measurement
In any state, the boot can be fixed by press the "start/stop" button for 2 seconds to turn it off. If the
device idle for 30 seconds, backlit closed, 60 seconds after the device has reached the "sleep" mode.
Press any button except/key can be reactivated backlit.In the condition of error and low electricity,
according to the key, open/equipment into "sleep" state immediately.
Low Battery Indicator
When measuring screen does not appear when the battery symbol, said the battery voltage is normal,
sphygmomanometer can work normally.
When the screen appear when measuring marks, said the battery voltage is low, please replace
the battery, prevent blood pressure monitor can not work normally.
Screen appears when measuring the flashing, and LCD screen will appear low voltage symbol
“LO”, sphygmomanometer automatic shutdown after about 5 s please immediately replace the battery,
ensure the sphygmomanometer can normal use.
Er5
LO