11. Alarm Removal Key
Alarm and Processing Control: when the O
2
concentration is higher than the set point and remains
constant for some time (to avoid false alarm for an abrupt increase of partial concentration), the O
2
Concentration Indicator Lamp is flashing with the alarm. Start both the left fan and the right fan to
enhance ventilation inside and outside. If the O
2
concentration remains unchanged, open the ICU door
to enhance ventilation. When the O
2
concentration is equal to or lower than the set point, switch off the
O
2
indicator lamp and close the alarm. If fans are not needed when other functions are working the
fans automatically stop.
1. Disinfection & Sterilization: after incubation or caring of one pet finishes treatment and care, it is
suggested to disinfect and sterilize this machine. Do this once every week when not in use.
2. Separate cleaning of the Humidifier
If the Humidifier fails to produce moisture, or the water spills out, take the following measures:
a) As shown in Figure 8.2.1-a,open the cover plate of the right Control Rack.
b) As shown in Figure 8.2.1-b and Figure 8.2.1-c,remove Water Level Sensors and Humidifier from the
Water Vessel. Take out the Water Vessel.
c) As shown in Figure 8.2.1-d, empty the Vessel. Brush off the mineral substance such as water scale
or verdigris on the Vessel, Water Level Sensors and Humidifier. If the filter of the Humidifier fails to
work, replace with a new filter.
d) Install Water Level Sensor and Humidifier, place the water vessel back and secure the cover plate.
Press "SET" key to remove disengage alarms. See fig. 7.11.1.
Humidifier
Water Level
Sensor
Highest Level
of Vessel
Water Vessel
Humidifier
Figure 8.2.1-b
Figure 8.2.1-a
8. Disinfec�on and Steriliza�on Control Part
9. Illumina�on Intensity Control
10. O
2 concentra�on monitoring
Press illumination adjustment key "+/-" to adjust
ten grades of illumination intensity. See fig. 7.9.1.
The highest is grade nine and the lowest is
closed. Under the Initial start-up status, it is
closed.
Key control: press "SET" key for five seconds to enter the parameters setting status. The humidity / O
2
display shows flashing digits of O
2
concentration set value. When the indicator lamp "P" (O
2
concentra-
tion) is flashing, it is able to adjust the set value of O
2
concentration alarm. Now press humidity "+/-"
key to adjust O
2
set value by ±1 stepping. The default factory setting is the alarm set point of 21%.
a) On: before starting the sterilization function, make sure that there is no pet in the ICU and the ICU
door is closed. Meanwhile, press "SET" key plus "sterilization" key to enter the disinfection mode. See
fig. 7.8.1.
b) Operation: under this mode, all other function keys are void except "SET" key and "sterilization" key.
Both temperature display and humidity display show "DISI" (abbreviated disinfection); under the
disinfection mode, switch on the UV disinfection lamp. There is violet light inside the ICU (with weak
acrid scent). Meanwhile, start the left fan and the right fan to effect ventilation inside.
c) Off: press "SET" key plus "disinfection" key at the same time again to switch off the UV disinfection
lamp, the disinfection indication lamp and the left fan to exit the disinfection mode and recover the
normal work mode. At this time all function keys are valid except the disinfection key.
Note: when the ICU is under the normal work mode, it is invalid to press "disinfection" key individually.
Figure 7.9.1
Figure 7.10.1
Figure 7.11.1
Hold for 5 seconds
Press “+/-” for se�ng the
valne for o
2
alarm
Press then remove the alarm
Set illumina�on
Press “+”, for brighter
Press “-”, for dimmer ten stage adjustable
14
15
Cleaning and Maintenance