Jorvet J1550 User Manual Download Page 8

5. Humidity Control Part

Control program:

a) When the setting temperature is higher than the measured temperature, start the PID precise

heating system and left/right fans. When the measured temperature is higher than the setting tempera-

ture, stop heating step by step.

b) When the measured temperature is higher than the setting temperature, start the PID precise

cooling system (need additionally purchased cooler) and the left fan to realize ventilation and air

exchange inside and outside. When the measured temperature is lower than the setting temperature,

stop cooling step by step and shut down fans after one-minute delay.

c) When there is no additionally connected cooler, try to place an ice tube inside the atomizing

chamber or at the sink as auxiliary cooling. The system will start the right fan and the left fan to realize 

ventilation and cooling.

    Indicator lamp control:

    Heating---the temperature indicator lamp is red

    Cooling---the temperature indicator lamp is blue

    Delay---the temperature indicator lamp is yellow

Before humidification, check if the water is enough 

in the internal water tank.

If you have an external humidifier and need high 

humidity in high temperature or dry environment, 

just install the external humidifier and start it before 

setting.

Key control: press "SET" key plus humidity "+/-" key 

at the same time to adjust the humidity set point, 

and regulate it by±0.1 stepping. See fig. 7.4.1. The 

factory setting is the humidity set point of 55%. It 

can be adjusted down to 40% and up to 65%.

Alarm function:

a) Alarm for beyond the maximum set point: the humidity and the upper dash are displayed

alternately every three seconds with alarm and flashing temperature indicator lamp.

b) Alarm for beyond the minimum set point: the humidity and the lower dash are displayed alternately

every three seconds with alarm and flashing temperature indicator lamp.

c) Damaged humidity sensor: flashing "ERR" display and flashing humidity indicator lamp in the humid-

ity display.

d) Internal humidifier water shortage alarm (the water level sensor of the humidifier is disconnected):

in this case, if the humidifier is started, the humidity Indicator lamp is off with water shortage alarm, and 

the humidity display shows "LAC". Meanwhile, it starts the smart water shortage protection system.

Control program:

a) When the setting humidity is higher the measured humidity, start the humidifier and the right fan.

b) When the measured humidity reaches the setting humidity, stop humidifying.

c) When the measured humidity is higher than the setting humidity, start dehumidifying and the left fan; 

reversely start the right fan for air exhaust as auxiliary dehumidifying. If an additionally purchased

cooler is connected, start the cooling function for dehumidifying.

Indicator lamp control:

The humidity indicator lamp is always on at the time of humidifying.

The humidity indicator lamp is flashing at the time of dehumidifying.

 Figure 7.5.1

Hold

6. Anion Control Part

7. Medical Atomiza�on Control

Press "anion" key to switch it ON; and press it 

again to switch it OFF.

When the anion is open, the anion indicator lamp 

is on; when the anion is closed, the anion indicator 

lamp is off. See fig. 7.6.1.

The atomizing volume of the medical atomizer has two grades, measuring by time and quantity. Press 

"medical atomization" key to start the atomizing function (it is the higher grade with the atomizing 

volume of 0.375ml/min and the auto-timing of 10 minutes). Press "atomizing volume regulation" key to 

switch between the higher grade and the lower grade of atomizing volume. The lower grade is 

0.5ml/min with the auto-timing of 20 minutes. When the medical atomization is started, the atomization 

indicator lamp is flashing quickly (correspondent to the higher grade). Press "regulation" key to adjust 

it to the lower grade at the time of which the indicator lamp is slowly flashing. Press "medical atomiza-

tion" key again to stop atomization, and the atomization indicator lamp is off.

a) Press "atomization" key to start the higher grade; press atomizing volume regulation key "+". To

adjust it to the lower grade, press atomizing volume regulation key "-". See fig. 7.7.1.

b) Press "atomization" key and get into the medical atomization status (higher grade) during which the 

humidity display and control are shielded. The humidity display shows accumulated atomization time 

counting (in seconds). After atomization and 10 sec delay, it automatically exits the atomization mode 

and reverts to the humidity control mode.

c) During atomization, press "atomization" key to manually stop the atomization mode and revert to the 

humidity control mode.

d) In the atomization mode, the humidity control is not effective. Therefore, when the humidity exceeds 

the maximum set point, there is no alarm. If the humidity still exceeds the maximum set point in three

minutes after it exits (closes) the medical atomization mode, the alarm is started.

Atomizer water shortage alarm: when the water tank or the medicine cup of the atomizer is short of

water, stop atomization. The humidity display shows "ELAC" and the atomization lamp is off with water 

shortage alarm. Press "atomization" key again to enter the "atomizer off" status and recover humidity

display.

 Figure 7.6.1

 Figure 7.7.1

 Figure 7.8.1

Hold press once message come out

Hold and press again for shut down

Press once for start

Press again to shut down

Press “+”, for 10 minutes pa�ern

Press “-”, for 20 minutes pa�ern

Press ones, start to 

nebuilizer, default 

state is for 10 ninute, 

press again for shut down

12

13

Summary of Contents for J1550

Page 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Page 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Page 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Page 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Page 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Page 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Page 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Page 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Page 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Page 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Page 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Page 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Reviews: