Jorvet J1550 User Manual Download Page 7

2. Temperature Display

3. Humidity/O

2

 Concentra�on Display

4. Temperature Control Part

Fahrenheit/centigrade can be switched and 

displayed with a toggle key. The default tempera-

ture is shown in 

fahrenheit

 and in two-digit integer 

and one decimal place after which is a letter 

showing special indication at the time of setting 

the temperature parameter. Switch the 

Fahrenheit / centigrade indicator lamp. When the 

indicator lamp displays "C", it means centigrade; 

when it displays "F", it means Fahrenheit. See fig. 

7.2.1.

Humidity / O

2

 concentration display can be 

switched with the humidity / O

concentration key. 

The default is humidity display in two-digit integer 

and one decimal place. O

2

 concentration is 

displayed in four-digit integer. Switch the humidity / 

O

2

 concentration indicator lamp. When the 

Indicator lamp is "HU", it means humidity; when the 

indicator lamp is“O

2

”, it means O

concentration. 

See fig. 7.3.1.

Key control: press "SET" key plus temperature " + / - " key at the same time to adjust the temperature 

set point, and regulate it by ± 0.1 stepping. See fig. 7.4.1.The factory setting is the set point of 

68

°

F

It can be adjusted down to 

59

°

F

 and up to 

104

°

F

;

Smart alarm function:

Alarm for beyond the maximum set point: the temperature and the upper dash are displayed alternate-

ly every three seconds with alarm and flashing temperature indicator lamp.

Alarm for beyond the minimum set point: the 

temperature and the lower dash are displayed 

alternately every three seconds with alarm and 

flashing temperature Indicator lamp.

Damaged temperature sensor: flashing "ERR" 

display and flashing temperature indicator lamp in 

the temperature display.

 Figure 7.4.1

 Figure 7.3.1

 Figure 7.2.1

Hold

10

11

7. ICU Illumina�on Func�on---- Crea�ng Comfortable Therapeu�c Environment

8. Se�ng of Security Mechanism---- No Worries in Use

9. Human Oriented Structural Design ---- Reasonable Opera�on, Clean and Convenient

1.Start-up and Shutdown

The 

JorVet  J1550  ICU  Unit

  is  equipped  with  a  precise  oxygen  concentration  monitor,  which 

can  monitor  changes  in  oxygen  concentration  in  the  ICU.  When  being  hypoxic  state  that 
concentration  falls  beyond  the  standard  value  (exceeds  the  set  value),  and  cannot  exceed  the  set 
value for three minutes, It will trigger the alarm system and start the intelligent ventilation system at 
the same time, being ventilated with the outside air and thus improve the air quality of ICU, 
which becomes the most important life defense line for pets.

JorVet  J1550  ICU  Unit

  uses  ten-level  LED  lights  adjustment,  which  can  simulate  various  scenes 

similar to parent, forest oxygen bar, warm home, forest dawn and dusk twilight, etc., providing a cozy 

and conformable therapeutic environment for pets.

JorVet  J1550  ICU  Unit

 adopts intelligent alarm system, and is equipped with 

corresponding fault display function, such as extreme temperature and excessive moisture 

alarm, lack-of-water alarm and main components fault warning lights.

JorVet J1550 ICU Unit

 is equipped with intelligent switch power keys, dividing into three controls, i.e. 

power-on, start and power-off. This design is energy saving and safer to use.

All equipment components of 

JorVet J1550 ICU Unit

 adopt integrated design. All compo-nents  are 

integrated in a control frame, and completely separated from the control circuit. Only by 

disassembling the control rack, the whole machine can be disassembled to clean various 

compo-nents, including fan, medical devices, heat sink and filter etc. The operation is extremely 

convenient, easy for troubleshooting, maintenance and equipment cleaning.

JorVet J1550 ICU Unit

 is equipped with two medical device import Interface channels, thus 

external medical equipment (such as oxygen machine) can be connected into the ICU, 

meanwhile, the side access port adopts medical valve importing design, being convenient for 

the importing of pets injection and infusion needles and relative sealing and fixing.

a)

One end of the power cord is plugged into the

terminal board at the bottom support of the right side of 

the machine, and the other end into the socket of 

110

0V / 

6

0Hz with reliable grounding. See fig. 7.1.2.

b)

The power switch is located at the batter brace of

the left side of the machine. See fig.7.1.2. It is the

on/off switch of 

JorVet ICU Unit

.

 

When it is locked,

all other keys are void; when it is

 

unlocked, all other 

keys are valid.

c)

The lock indicator lamp is at the on/off switch. 

When it is on, it means unlocked; when it is off, it

means

 

locked. 

 Figure 7.1.2

Power ON/OFF

Power Interface

Product Operation Instructions

Summary of Contents for J1550

Page 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Page 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Page 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Page 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Page 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Page 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Page 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Page 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Page 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Page 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Page 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Page 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Reviews: