2. Temperature Display
3. Humidity/O
2
Concentra�on Display
4. Temperature Control Part
Fahrenheit/centigrade can be switched and
displayed with a toggle key. The default tempera-
ture is shown in
fahrenheit
and in two-digit integer
and one decimal place after which is a letter
showing special indication at the time of setting
the temperature parameter. Switch the
Fahrenheit / centigrade indicator lamp. When the
indicator lamp displays "C", it means centigrade;
when it displays "F", it means Fahrenheit. See fig.
7.2.1.
Humidity / O
2
concentration display can be
switched with the humidity / O
2
concentration key.
The default is humidity display in two-digit integer
and one decimal place. O
2
concentration is
displayed in four-digit integer. Switch the humidity /
O
2
concentration indicator lamp. When the
Indicator lamp is "HU", it means humidity; when the
indicator lamp is“O
2
”, it means O
2
concentration.
See fig. 7.3.1.
Key control: press "SET" key plus temperature " + / - " key at the same time to adjust the temperature
set point, and regulate it by ± 0.1 stepping. See fig. 7.4.1.The factory setting is the set point of
68
°
F
.
It can be adjusted down to
59
°
F
and up to
104
°
F
;
Smart alarm function:
Alarm for beyond the maximum set point: the temperature and the upper dash are displayed alternate-
ly every three seconds with alarm and flashing temperature indicator lamp.
Alarm for beyond the minimum set point: the
temperature and the lower dash are displayed
alternately every three seconds with alarm and
flashing temperature Indicator lamp.
Damaged temperature sensor: flashing "ERR"
display and flashing temperature indicator lamp in
the temperature display.
Figure 7.4.1
Figure 7.3.1
Figure 7.2.1
Hold
10
11
7. ICU Illumina�on Func�on---- Crea�ng Comfortable Therapeu�c Environment
8. Se�ng of Security Mechanism---- No Worries in Use
9. Human Oriented Structural Design ---- Reasonable Opera�on, Clean and Convenient
1.Start-up and Shutdown
The
JorVet J1550 ICU Unit
is equipped with a precise oxygen concentration monitor, which
can monitor changes in oxygen concentration in the ICU. When being hypoxic state that
concentration falls beyond the standard value (exceeds the set value), and cannot exceed the set
value for three minutes, It will trigger the alarm system and start the intelligent ventilation system at
the same time, being ventilated with the outside air and thus improve the air quality of ICU,
which becomes the most important life defense line for pets.
JorVet J1550 ICU Unit
uses ten-level LED lights adjustment, which can simulate various scenes
similar to parent, forest oxygen bar, warm home, forest dawn and dusk twilight, etc., providing a cozy
and conformable therapeutic environment for pets.
JorVet J1550 ICU Unit
adopts intelligent alarm system, and is equipped with
corresponding fault display function, such as extreme temperature and excessive moisture
alarm, lack-of-water alarm and main components fault warning lights.
JorVet J1550 ICU Unit
is equipped with intelligent switch power keys, dividing into three controls, i.e.
power-on, start and power-off. This design is energy saving and safer to use.
All equipment components of
JorVet J1550 ICU Unit
adopt integrated design. All compo-nents are
integrated in a control frame, and completely separated from the control circuit. Only by
disassembling the control rack, the whole machine can be disassembled to clean various
compo-nents, including fan, medical devices, heat sink and filter etc. The operation is extremely
convenient, easy for troubleshooting, maintenance and equipment cleaning.
JorVet J1550 ICU Unit
is equipped with two medical device import Interface channels, thus
external medical equipment (such as oxygen machine) can be connected into the ICU,
meanwhile, the side access port adopts medical valve importing design, being convenient for
the importing of pets injection and infusion needles and relative sealing and fixing.
a)
One end of the power cord is plugged into the
terminal board at the bottom support of the right side of
the machine, and the other end into the socket of
110
0V /
6
0Hz with reliable grounding. See fig. 7.1.2.
b)
The power switch is located at the batter brace of
the left side of the machine. See fig.7.1.2. It is the
on/off switch of
JorVet ICU Unit
.
When it is locked,
all other keys are void; when it is
unlocked, all other
keys are valid.
c)
The lock indicator lamp is at the on/off switch.
When it is on, it means unlocked; when it is off, it
means
locked.
Figure 7.1.2
Power ON/OFF
Power Interface
Product Operation Instructions