This External humidifier is helping the internal humidifier in our ICU to reach desired RH level more 

easily, especially in extremely dry environment or under high temperature. The power plug should be 

plugged into two flat pins receptacle at the rear left corner on the back of the ICU unit, it will be 

regulated by the ICU motherboard with special programs installed. (Note: Do not confuse this socket 

with another socket for nebulizer, which is next to this socket for humidifier.) Remove the lid of the air 

inlet at the back of the ICU, expose the portal, and connect the flexible hose from this humidifier to 

the ICU air inlet. The end with large diameter is to the back portal of ICU and the end with small 

diameter it to the top portal of the humidifier.

After the unit is plugged into either ICU power supply outlet or wall mains, you can start the machine.

 

Rotate clockwise the controller to switch on the unit and the blue indicator lamp will be on. 

The unit starts working. Rotate the controller knob to adjust the volume of the vapor. Fog can be seen 

from the outlet port.

If a warm vapor is needed, press the button below the power button to switch between the default cool 

mist mode and warm mist mode.

Remarks: the “warm mist mode” will be needed when the ICU unit is set at “high tempe high 

humidity”, like greater than 28

 + 60%RH, the warm mist will definitely help humidifying more quickly.

 

Note

When LAC alarm can be seen on the humidity display panel, the external and internal 

humidifiers will both stop working. Fill water in the internal water tank to the maximum level and both of 

the humidifiers will start to work.

In certain cases, such as avian or other exotics, a desired environment of higher humidity and 
temperature is preferred. In order to achieve these levels you must have increased temperature and 
increased humidity. For instance, at 100F the humidity can not go beyond 50% relative humidity 
without the external humidifier.
The external humidifier is not necessary when using the unit on standard mammalian patients like 
canine and feline. 

4. External Humidifier Installa�on

 (optional J1550D1)

04

05

The patented Energy Saving Low Noise Compressor (less than 25 decibels) generates high speed air 

jet from filtered clean air (by anti-bacterial cotton), the air stream disperses the medicine solution from 

liquid cup into tiny particles and form medicine aerosol. The average size of droplet could reach 2.5μ

m minus which could reach deep respiratory system like alveolus and fine bronchus. The medicine is 

nebulized evenly, no diluted needed, so there is no residues left almost and could achieve very high 

medicine utilization rate. This device is preciously made medical equipment with very low faulty rate 

and very good durability, could serve for 1000 -1500 hours as minimum lifespan.

Operation Instructions: 

The nebulizer is easy to operate, please follow the simple steps as follows.

1. Install the nebulizer on the left side of ICU. Insert the power plug into the socket on the ICU.

2. Screw off the cup lid and nozzle; pour medicine liquid into the nebulizer cup.

3. Screw on the cup lid and nozzle, and put the nebulizer back.

4. Insert the nozzle of Nebulizer into the port. Turn on the nebulizer by pressing the switch on the

nebulizer. And then press Nebulize button on control panel of ICU to switch on nebulization, the

medical fog will be sprayed through grid and get into the ICU chamber. There are two stages of

nebulization treatment timing, in which the basic time is 10 minutes. Once the nebulization process

started, it is 10 minutes countdown automatically; pressing the "-" button on the nebulize adjustment

button will turn it into 20 minutes mode, pressing the "+" will turn back to 10 minutes mode. Pressing

Nebulize button again on the main control panel when the nebulizing process is running will switch the 

nebulizer off.

a)

The treatment process should not exceed 20 minutes for every nebulizing, misuse would possibly

cause Pulmonary edema

.

b)

To prevent superinfection, please pay close attention to the sterilization of nebulizer, ambient air and 

all the related equipment, also disinfect animal mouth, nose and throat.

Mechanism of action

Animal respiratory system is an open system, where alveolus is the main air-exchanging place. The 

object sprayed out of the nebulizer is mainly the mixture of oxygen and medicine – oxygen-medicine 

aerosol. This kind of aerosol could improve oxygen metabolism and anoxic metabolism significantly; 

increase the oxygen reserve in organisms. The cavitation effect of compressed air could disperse the 

liquid medicine into gas phase and turn the liquid into fog form. The suspended particles will get deep 

into animal respiratory system and treat the focus of infection directly. The feature of air compressing 

nebulization technique is generating Big Fog Volume Fog with Ultra Small Particle Size (0.5 -5μm); 

the 

even  fog  could  go  very  deep  into  bronchus  and  the  particles  could  settle  down  in  deep 

respiratory system, achieves painless cure effect. 

Tips: 

a) The smaller the particles are, the better curative effect the nebulization could achieve.

b) Normally the fog with bigger particles generated from traditional ultrasonic nebulizer would be

easier to be observed, as the fog seems to be thicker.

c) The international of human medical nebulizer is the particles in fog need to be smaller than 5μm.

Specialized Medical Nebulizer for Animal – Be�er Cura�ve Effect

Air  Compressor  Nebulizer  uses  high  pressure  gas  jet  knockout  the  medicine  solution  into  tiny 

particles and carry them into deep respiratory system. The size of droplet will affect the settlement site 

of medical fog.

D

roplet Si

z

Se

lement Site 

1

-

5

μ

Fine 

B

ronc

h

us

,

 

A

lveolus 

5

-

20

μ

B

ronc

h

us 

20

-

40

μ

Upper 

R

espiratory Tract mainly

,

 Nose

,

 T

h

roat

,

 etc

.

 

Unscrew the tank cover

Add water

Take the water tank out of the base, take away the Valve-cup, and fill the water tank with clear water. 

Close the Valve-cup, and put the water tank on to the base carefully.

Nebulization

NOTE: NEVER RUN THE UNIT WHEN THERE IS NO WATER IN WATER 

BASIN. 

Usage Warning:

Summary of Contents for J1550

Page 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Page 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Page 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Page 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Page 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Page 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Page 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Page 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Page 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Page 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Page 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Page 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Reviews: