2
CD-Wxx-00-0 Series Wall Mount CO
2
Transmitter Installation Instructions
Surface Mounting
Use the following procedure to mount the CO
2
transmitter to a drywall surface:
1. Using a hole saw with a 1-3/8 in. blade, drill the
center hole in the surface where the sensor will be
mounted, then pull the wiring through the drilled
hole.
2. Using only the drywall mounting holes indicated in
Figure 1, mark the location of the two mounting
holes and drill the 1/8 in. holes.
Center
Hole
Drywall
Mounting
Hole
Drywall
Mounting
Hole
Wallbox
Mounting
Hole
Wallbox
Mounting
Hole
Figure 1: Transmitter Base
Spacer (2) for
3/8 in. (9.5 mm) and 1/2 in.
(12.7 mm) Thick Drywall
1-3/8 in.
(35 mm)
Hole
Drywall
Drywall
Spring
Clip (2)
Top View
Drywall
Spring
Clips (2)
1-3/8 in.
(35 mm)
Hole
Base
Two No. 6
Self-Tapping
Screws (Provided)
Figure 2: Surface Mounting to Drywall
3. Insert the spring clips as shown in Figure 2. (Use
the spacers for 3/8 or 1/2 in. drywall.)
Note:
Replacement spring clips, spacers, and
screws can be ordered separately as the
Drywall Mounting Kit (ACC-DWCLIP-0).
4. Pull the wiring through the base’s center hole.
5. Center the two holes in the base over the drilled
holes, and fasten the base to the drywall using the
No. 6 screws provided (Figure 1 and Figure 2).
!
CAUTION:
When mounting to drywall, use
the specified holes (see Figure 2) with the provided
spring clips. Other mounting may cause the screws
to damage the PCB or cause a short circuit.
IMPORTANT:
To prevent drywall damage,
do not overtighten the screws.
6. To install the PCB into the base, refer to Figure 3
and:
•
align the PCB over the latch pins on the base.
•
press down on the board until it snaps into
place and is secured by the latches.
7. Proceed to the
Wiring
section.