INSIGHT Lateral Access System Surgical Technique DePuy Synthes 17
Slide the reusable light (1) or disposable light (2) into the
light slots. Secure light underneath the provided hooks.
Connect the light cable to the light source. Turn on the
light source.
Precautions and Warnings
•
Do not bend fiber optic cables/lights under a radius
of 5 cm.
•
Do not apply pressure on the light cable/lights
using a sharp object.
•
Exchange reusable light/cables if it collects fluid
inside, appears broken or damaged.
•
Avoid damaging the fiber surfaces at the ends of
the light cable, as this will reduce the light output
level.
•
Do not use higher wattage than indicated for
the light cables and reusable light (300 W).
•
Depending on light source, temperature of the
light, cables and/or adapters may exceed 43°C.
Therefore avoid contact to user, patient, tempera
ture-sensitive objects and flammable materials
such as textiles (curtains) or near cotton swabs or
pads that have been soaked with flammable fluids
with these parts.
•
The reusable light/disposable light should only be
used with the associated light cables.
•
Never leave the light system unattended when light
is being transmitted from a light source.
•
Never look directly into the highly intense light
since this could cause severe injuries to the eyes.
•
The light instruments containing fiber optics
should not be ultrasonically cleaned.