Forward/Backward Retraction Options
Retract Forward
If forward retraction is desired
(A),
attach the connector
to the
third blade holder
connection point.
Subsequent clockwise rotation of the RETRACT nut
results in forward retraction of the retractor body (with
left and right blade holders) relative to the fixed third
blade holder.
or
Retract Backward
If backward retraction is desired
(B),
attach the connec-
tor to either one of the attachment points on the
retractor body.
Subsequent clockwise rotation of the RETRACT nut
results in backward retraction of the third blade holder
relative to a fixed retractor body (with left and right
blade holder).
Precaution:
Do not place any accessories before
retraction.