Problem

Possible Cause

What to Do

Bed movement with all 
casters locked.

Braking mechanisms may be worn 
on one or more casters.

1.

Remove bed from use and contact 
Joerns Healthcare Technical Support 

at 800-966-6662.

1.

Scale System display  
is blank after On button 
is pushed.

Batteries are dead.
Unit is defective.

1.

2.

Replace all six batteries

Contact Joerns Healthcare Technical 
Support at 800-966-6662.

1.

2.

Scale System gives 
suspect readings.

Scale System cable  
is disconnected.
Scale System may need calibration.
5th Wheel is engaged.

1.

2.
3.

Reattach cable. 

Refer to Page 13 of this Manual.

Disengage 5th Wheel.

1.

2.
3.

Mattress Support 
Platform sections do  
not extend.

Latch handles not fully disengaged. 

 

Binding

1.

2.

While actuating the latch handle, 

try to extend the mattress support 
platform extension. Refer to page 9 
of this manual.
Wider support platform sections may 
require two hands to apply equal pull 

force. Contact Joerns Healthcare 
Technical Support at 800-966-6662.

1.

2.

Mattress Support 
Platform sections do  
not articulate.

Pendant Control operation.

 

Selective function lockout may have 
been activated.
Unit may be defective or damaged.

1.

2.

3.

Refer to Pendant Control 
Troubleshooting on page 18.
Refer to page 8 of this manual, 

Selective Function Lockout.
Remove bed from use and contact 
Joerns Healthcare Technical Support 

at 800-966-6662.

1.

2.

3.

Actuator piston rod will 
only move inwards and 
not outwards.

Unit may be defective or damaged. 

Possible synchronization problems.

1.

Remove bed from use and contact 
Joerns Healthcare Technical Support 

at 800-966-6662.

1.

Actuator motor makes 

sound, but piston rod 

does not move. 

Unit may be defective or damaged.

1.

Remove bed from use and contact 
Joerns Healthcare Technical Support 

at 800-966-6662.

1.

Actuator runs too  
slowly or does not  
give full force.

Unit is receiving insufficient power 
or, is defective. Not plugged in and 

running on batteries

1.

Remove bed from use and contact 
Joerns Healthcare Technical Support 

at 800-966-6662.

1.

Summary of Contents for Bari10A

Page 1: ...Joerns Healthcare User Service Manual Joerns Model Bari10A Bariatric Bed leading bed evolution To avoid injury read user s manual before using bed...

Page 2: ...not be placed in an oxygen enriched environment such as an oxygen tent Use of electrical circuits in an oxygen enriched environment could result in a fire hazard 2 Warning Possible Shock Hazard Unplu...

Page 3: ...maneuverability performance push the bed from the head end To avoid injury to the resident patient and or user do not attempt to move the bed laterally with the steer mode engaged The steer wheel can...

Page 4: ...f long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance...

Page 5: ...ra 8 length for taller residents patients 18 3 4 to 33 3 4 Travel Range Integrated Scale System for resident patient weight monitoring Head and Foot End Side Rails Assist Devices Side Rails on Head an...

Page 6: ...Lock for all functions to keep bed from accidentally operating using battery backup Note To avoid electric shock always plug bed into properly grounded hospital grade receptacle Actuator battery backu...

Page 7: ...port Platform Release Lever Up Hold Emergency Release handle Up until Mattress Support Platform lowers to desired position Release handle if you need to stop descent 1 1 1 2 7 Figure 4 Bari Selective...

Page 8: ...s rotate switch until it clicks into position Side Rails Assist Devices Bari10A has one Head end Side Rail and one Foot end Side Rail on both sides of bed Figure 7 Side Rails are attached and operated...

Page 9: ...owest position with the siderails fully raised and locked when moving the bed with a resident patient in it For best maneuverability performance push the bed from the head end Warning 5th wheel must b...

Page 10: ...p process Slide the Frame Extender out first by pulling out the latch knob on the side of the bed frame Figure 14 Begin carefully sliding the extender out and release the latch knob as soon as the ext...

Page 11: ...t to lock all four casters individually to prevent unintended bed movement Always plug bed back into power outlet so that it is continuously charging Accessories Mattress Bari10A Bed may be equipped w...

Page 12: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 12 F Figure 18...

Page 13: ...ng Conditions Ambient Temperature 20 C to 30 C Relative Humidity 10 to 95 Non Condensing Storage and Shipping Conditions Ambient Temperature 30 C to 50 C Relative Humidity 10 to 95 Non Condensing 1 2...

Page 14: ...will lose all stored resident patient s weight history Pressing Weight Mode button will switch back to the Weight Mode allowing you to view resident patient s total weight again Caution If you do not...

Page 15: ...12 months or as your maintenance policy requires Electrostatic Discharge ESD Integrated circuits and semiconductors on printed circuit boards may be damaged by electrostatic discharge ESD Be sure to...

Page 16: ...teries 1 2 3 1 2 3 4 Battery Replacement If Scale System batteries need replacement display will indicate CHANGE BATTERIES Remove Battery Cover from Foot Panel to access Scale System batteries Figure...

Page 17: ...Use Super Lube synthetic grease or equivalent Oiling Points Keep pins and bolts oiled sample indicated by arrows in Figure 24 Oil hinge points sample indicated by arrows in Figure 25 Grease Points For...

Page 18: ...d may be damaged Check cord for damage If cord is damaged discontinue bed usage until cord is replaced 1 2 3 4 If Adjustment Only partially occurs stops and beeps Battery may be low Plug bed into powe...

Page 19: ...ctions may require two hands to apply equal pull force Contact Joerns Healthcare Technical Support at 800 966 6662 1 2 Mattress Support Platform sections do not articulate Pendant Control operation Se...

Page 20: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 20 Appendix A Actuator Wiring Diagram...

Page 21: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model Appendix B Scale System Wiring Diagram 21...

Page 22: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 22 Notes...

Page 23: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 23 Notes...

Page 24: ...mal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to...

Reviews: