Remove weight and re-zero. Repeat steps #4, 
#5 and #6, until results are satisfactory. 

When adjustments are completed:

 

a.  Press 

on

 button to 

abort

 your changes    

 

 

and exit the CAL mode, 

(or)

 

 

b.  Press 

Freeze

 button to 

save

 all changes  

 

 

and exit CAL mode. 

Calibration Tolerance Table

Low Limit

Applied Load (lbs)

High Limit

248.7

250

251.4

497.4

500

502.6

746.2

750

753.9

994.9

1000

1005.1

Scale System Care  

and Maintenance

Cleaning 

Mild soap/water recommended for general 
cleaning/disinfecting. 
Exercise caution when cleaning LED 
electronics display window. 

 

Storage 

Remove batteries if storing this equipment for 
periods longer than three months. 

7.

8.

1.

2.

Scale System Calibration 

(Figure 22) 

Periodic Calibration Check 

Important:

 Following calibration procedure 

should only be performed by qualified 

maintenance personnel. 
Load cell has no user serviceable components. 

Do not 

tamper with for any reason. 

Re-calibration is generally not required, but 
should be verified periodically to ensure 

accuracy. Joerns Healthcare recommends 

checking calibration check once every 12 
months, or as your maintenance policy requires. 

Electrostatic Discharge (ESD)

Integrated circuits and semiconductors on printed 
circuit boards may be damaged by electrostatic 

discharge (ESD). Be sure to use proper handling 

precautions at all times. 

Calibration Procedure 

Note: 

Make sure nothing is in contact with Scale 

System during this procedure. Be sure to remove 
your hands from Scale System when noting  
displayed calibration results. 

Disengage 5th Wheel if equipped with optional 
steer pedal.
Press 

Freeze/Resume

 button. 

When display indicates “FREEZE = (weight)”, 

press and hold 

Lb/Kg

 button. Display will read 

“CAL” as right hand digit counts down to enter 
CAL MODE. 
When CAL MODE is shown, press 

Zero

 button 

to zero the display. 

Place known calibrated weight onto weighing 
surface and compare it to displayed reading. 
Press 

Weight Mode

 button for (+) corrections 

and 

Weight Change Mode

 button for (-) 

corrections to displayed weight. Displayed value 

should be within 0.5% of calibrated weight, plus 

or minus one digit. 

1.

2.

3.

1.

2.
3.

4.

5.

6.

Figure 23

15

Figure 22

0.0

zero

Freeze

Resume

Weight

Change

Mode

On

Note:

Material - 

7 mil GLOSS POLYESTER - see finish notes below
Emboss Tool - 0.015”- 0.010” pillow emboss (for metal domes) 

6 Buttons up to edge of and including the white region (ellipses) -

Note - Two Hidden buttons - (flat non tactile) see sheet 2of 2
Selective Hardcoat to be Velvet except in region of 6 keys (matching emboss) 

and Display Window

Key embosses (six) to be Gloss 

Adhesive Mask - selective adhesive tool

No adhesive behind clear window

Adhesive 3M part no 7959MP

Display window to be Gloss 

See sheet 2 of 2 for dimensional layout, circuit layout, tail layout

and more specifications 

Edge Bleed to be added

Selective adhesive die

on its own layer - 

PMS 2715 C

Purple

PMS 425 C

Dark Gray

PMS 360 C

Green

PMS 429 C

Light Gray 

OPAQUE

WHITE

Screened  colors 

Lb/Kg

1

.8

6

"

7.42"

Weight

Mode

.06 R. TYP. 4

NOTE - SEE SHEET 2 OF 2 

FOR TWO HIDDEN KEYS

Rev

Part #:

Page 1 of 2

216-2268

Joerns Bari 10A Bed

Scale Membrane Keypad

INIT

Scott Technology, LLC
1 Jacks Hill Road
Oxford, CT 06478

Scott

Init ________/Date_________

Nexus

Init ________/Date________

Date

ECO

JP

00430

B

03/27/07

Scott

Init ________/Date_________

03/20/07

JP

-

A

Summary of Contents for Bari10A

Page 1: ...Joerns Healthcare User Service Manual Joerns Model Bari10A Bariatric Bed leading bed evolution To avoid injury read user s manual before using bed...

Page 2: ...not be placed in an oxygen enriched environment such as an oxygen tent Use of electrical circuits in an oxygen enriched environment could result in a fire hazard 2 Warning Possible Shock Hazard Unplu...

Page 3: ...maneuverability performance push the bed from the head end To avoid injury to the resident patient and or user do not attempt to move the bed laterally with the steer mode engaged The steer wheel can...

Page 4: ...f long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance...

Page 5: ...ra 8 length for taller residents patients 18 3 4 to 33 3 4 Travel Range Integrated Scale System for resident patient weight monitoring Head and Foot End Side Rails Assist Devices Side Rails on Head an...

Page 6: ...Lock for all functions to keep bed from accidentally operating using battery backup Note To avoid electric shock always plug bed into properly grounded hospital grade receptacle Actuator battery backu...

Page 7: ...port Platform Release Lever Up Hold Emergency Release handle Up until Mattress Support Platform lowers to desired position Release handle if you need to stop descent 1 1 1 2 7 Figure 4 Bari Selective...

Page 8: ...s rotate switch until it clicks into position Side Rails Assist Devices Bari10A has one Head end Side Rail and one Foot end Side Rail on both sides of bed Figure 7 Side Rails are attached and operated...

Page 9: ...owest position with the siderails fully raised and locked when moving the bed with a resident patient in it For best maneuverability performance push the bed from the head end Warning 5th wheel must b...

Page 10: ...p process Slide the Frame Extender out first by pulling out the latch knob on the side of the bed frame Figure 14 Begin carefully sliding the extender out and release the latch knob as soon as the ext...

Page 11: ...t to lock all four casters individually to prevent unintended bed movement Always plug bed back into power outlet so that it is continuously charging Accessories Mattress Bari10A Bed may be equipped w...

Page 12: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 12 F Figure 18...

Page 13: ...ng Conditions Ambient Temperature 20 C to 30 C Relative Humidity 10 to 95 Non Condensing Storage and Shipping Conditions Ambient Temperature 30 C to 50 C Relative Humidity 10 to 95 Non Condensing 1 2...

Page 14: ...will lose all stored resident patient s weight history Pressing Weight Mode button will switch back to the Weight Mode allowing you to view resident patient s total weight again Caution If you do not...

Page 15: ...12 months or as your maintenance policy requires Electrostatic Discharge ESD Integrated circuits and semiconductors on printed circuit boards may be damaged by electrostatic discharge ESD Be sure to...

Page 16: ...teries 1 2 3 1 2 3 4 Battery Replacement If Scale System batteries need replacement display will indicate CHANGE BATTERIES Remove Battery Cover from Foot Panel to access Scale System batteries Figure...

Page 17: ...Use Super Lube synthetic grease or equivalent Oiling Points Keep pins and bolts oiled sample indicated by arrows in Figure 24 Oil hinge points sample indicated by arrows in Figure 25 Grease Points For...

Page 18: ...d may be damaged Check cord for damage If cord is damaged discontinue bed usage until cord is replaced 1 2 3 4 If Adjustment Only partially occurs stops and beeps Battery may be low Plug bed into powe...

Page 19: ...ctions may require two hands to apply equal pull force Contact Joerns Healthcare Technical Support at 800 966 6662 1 2 Mattress Support Platform sections do not articulate Pendant Control operation Se...

Page 20: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 20 Appendix A Actuator Wiring Diagram...

Page 21: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model Appendix B Scale System Wiring Diagram 21...

Page 22: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 22 Notes...

Page 23: ...www joerns com info joerns com Joerns Bariatric Beds Bari10A Model 23 Notes...

Page 24: ...mal use These parts such as DC batteries are not covered under the one year warranty but are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to...

Reviews: