Remove weight and re-zero. Repeat steps #4,
#5 and #6, until results are satisfactory.
When adjustments are completed:
a. Press
on
button to
abort
your changes
and exit the CAL mode,
(or)
b. Press
Freeze
button to
save
all changes
and exit CAL mode.
Calibration Tolerance Table
Low Limit
Applied Load (lbs)
High Limit
248.7
250
251.4
497.4
500
502.6
746.2
750
753.9
994.9
1000
1005.1
Scale System Care
and Maintenance
Cleaning
Mild soap/water recommended for general
cleaning/disinfecting.
Exercise caution when cleaning LED
electronics display window.
Storage
Remove batteries if storing this equipment for
periods longer than three months.
7.
8.
1.
2.
Scale System Calibration
(Figure 22)
Periodic Calibration Check
Important:
Following calibration procedure
should only be performed by qualified
maintenance personnel.
Load cell has no user serviceable components.
Do not
tamper with for any reason.
Re-calibration is generally not required, but
should be verified periodically to ensure
accuracy. Joerns Healthcare recommends
checking calibration check once every 12
months, or as your maintenance policy requires.
Electrostatic Discharge (ESD)
Integrated circuits and semiconductors on printed
circuit boards may be damaged by electrostatic
discharge (ESD). Be sure to use proper handling
precautions at all times.
Calibration Procedure
Note:
Make sure nothing is in contact with Scale
System during this procedure. Be sure to remove
your hands from Scale System when noting
displayed calibration results.
Disengage 5th Wheel if equipped with optional
steer pedal.
Press
Freeze/Resume
button.
When display indicates “FREEZE = (weight)”,
press and hold
Lb/Kg
button. Display will read
“CAL” as right hand digit counts down to enter
CAL MODE.
When CAL MODE is shown, press
Zero
button
to zero the display.
Place known calibrated weight onto weighing
surface and compare it to displayed reading.
Press
Weight Mode
button for (+) corrections
and
Weight Change Mode
button for (-)
corrections to displayed weight. Displayed value
should be within 0.5% of calibrated weight, plus
or minus one digit.
1.
2.
3.
1.
2.
3.
4.
5.
6.
Figure 23
15
Figure 22
0.0
zero
Freeze
Resume
Weight
Change
Mode
On
Note:
Material -
7 mil GLOSS POLYESTER - see finish notes below
Emboss Tool - 0.015”- 0.010” pillow emboss (for metal domes)
6 Buttons up to edge of and including the white region (ellipses) -
Note - Two Hidden buttons - (flat non tactile) see sheet 2of 2
Selective Hardcoat to be Velvet except in region of 6 keys (matching emboss)
and Display Window
Key embosses (six) to be Gloss
Adhesive Mask - selective adhesive tool
No adhesive behind clear window
Adhesive 3M part no 7959MP
Display window to be Gloss
See sheet 2 of 2 for dimensional layout, circuit layout, tail layout
and more specifications
Edge Bleed to be added
Selective adhesive die
on its own layer -
PMS 2715 C
Purple
PMS 425 C
Dark Gray
PMS 360 C
Green
PMS 429 C
Light Gray
OPAQUE
WHITE
Screened colors
Lb/Kg
1
.8
6
"
7.42"
Weight
Mode
.06 R. TYP. 4
NOTE - SEE SHEET 2 OF 2
FOR TWO HIDDEN KEYS
Rev
Part #:
Page 1 of 2
216-2268
Joerns Bari 10A Bed
Scale Membrane Keypad
INIT
Scott Technology, LLC
1 Jacks Hill Road
Oxford, CT 06478
Scott
Init ________/Date_________
Nexus
Init ________/Date________
Date
ECO
JP
00430
B
03/27/07
Scott
Init ________/Date_________
03/20/07
JP
-
A