Bari10A Bed Troubleshooting
Pendant Control Troubleshooting
Problem
Solution
If adjustment does not occur when
Pendant button is pushed
Adjustment may be locked. Check Selective Funtion Lockout.
Adjustment may already be at maximum level.
Bed is not plugged into wall power outlet and battery has
discharged. Plug bed into wall outlet.
Pendant cord may be damaged. Check cord for damage. If cord is
damaged, discontinue bed usage until cord is replaced.
1.
2.
3.
4.
If Adjustment Only partially occurs,
stops and beeps
Battery may be low. Plug bed into power source.
Bed may be overloaded or constrained from moving. Determine
load and check for obstructions.
If beeping occurs while raising or lowering bed, Hi-Low Actuators
may have lost synchronization. Lower bed to lowest position.
Release Bed Down button. Press and hold Bed Down button for
five seconds to recalibrate synchronization. If problems persist,
contact Joerns Healthcare Technical Support at 800-966-6662.