PAGE 26 of 53
1.
Ensure the probe holder is in latched, upper position
(Fig. 95)
.
2.
Using the supplied 3 mm hex driver loosen the transverse adjustment screw
(Fig. 99)
and rotate the yoke about the vertical shaft achieving the desired
angle.
3.
Tighten the transverse adjustment screw
(Fig. 100)
.
To return the transverse adjustment to neutral
(90°)
. The probe holder must be in
the latched, upper position
(Fig. 95)
. Rotate the yoke until the stop post contacts
the base of the probe holder
(Fig. 101)
. Then tighten the transverse adjustment
screw.
4.15.4. Probe Holder Longitudinal Adjustment
To adjust the probe holder’s vertical angle for longitudinal scanning, follow these
steps:
1.
Ensure the probe holder is in latched, upper position
(Fig. 95)
.
2.
Using the supplied 3 mm hex driver
(Fig. 27)
, loosen the longitudinal
adjustment screw
(Fig. 102)
.
3.
Rotate the main body of the probe holder until it is at the desired angle.
4.
Tighten the longitudinal adjustment screw
(Fig. 103)
.
5.
To return the longitudinal adjustment to neutral
(90°)
. Line up the longitudinal
adjustment indicator markers
(Fig. 104)
.
Fig. 102
- Loosen 3 mm screw
Fig. 103
- Rotate to position
Fig. 104
- Line up markers