PAGE 17 of 53
CE0120 Rev 05
To adjust the probe holder’s force, follow these steps:
NOTE:
Do not perform this operation on scanning surface.
1.
Ensure the probe holder is in the upright latched position
(Fig. 61)
.
2.
Lift probe holder slightly and press the latch button
(Fig. 65)
to release the
probe holder the full 45° degrees
.
3.
Insert the short arm of a 3 mm hex wrench into the 3 mm slot
(Fig. 66)
.
4.
Place the 2 mm hex wrench into the force adjustment screw
(Fig. 67)
.
5.
Lightly press the long arm of the 3 mm hex wrench down. Using the 2 mm
hex wrench, loosen the force adjustment screw but do not remove it
(Fig. 68)
.
6.
Gently apply pressure on
the long leg of the 3 mm
hex wrench until the force
adjustment marker lines
up with the desired spring
tension. While keeping the
markers in line, tighten the
force adjustment screw
(Fig. 69)
.
Fig. 65
- Lift slightly and press latch
Fig. 66
- Unlatched position
Fig. 67
- Insert hex wrenches
Fig. 68
- Press 3 mm hex wrench down
Heavy
Medium
Light
Force Adj. Marker
Heavy
Medium
Light
Force Adj. Marker
Fig. 69
- Choose desired tension