34
saved, press the "Reset" switch to turn off the output.
9.13 Correction of AC voltage to ground
Press the "SET" key, the program enters the ground resistance AC voltage correction mode, and the
display will show:
Connect a standard voltmeter capable of measuring 10VAC to the test end of the instrument, and
then press the "Start" key. At this time, the calibration program of the instrument will
automatically output a voltage of about 6VAC, and the display will show:
Please use the "+" or "-" key on the panel to input the standard voltage into the calibration
program, press the "+" key to increase the number, and the "-" key to decrease the number.
Please confirm the number is correct before pressing the "EXIT" key to close the output and
save the input data. If the data is not saved, press the "Reset" switch to turn off the output.
9.14 Correction of ground resistance AC current
Press the "SET" key, the program enters the ground resistance AC current correction mode, and
the display will show:
Connect a standard ammeter capable of measuring 30AAC to the test end of the instrument, and
then press the "Start" key. At this time, the calibration program of the instrument will
automatically output a current of about 25AAC, and the display will show:
Please use the "+" or "-" key on the panel to input the standard value of the standard ammeter
into the calibration program, the unit is A. Please confirm the number is correct before pressing
the "EXIT" key to close the output and save the input data. If the data is not saved, press the
"Reset" switch to turn off the output.
CAL GND V = 6
.
000V
<TEST> to Calibrate
CAL GND I = 25.00A
<TEST> to Calibrate
CAL GND V = X
.
XXXV
<EXIT> to Save
CAL GND I = XX
.
XXA
<EXIT> to Save