Please use the "+" or "-" key on the panel to input the standard voltage into the calibration
program, and its unit is V. Please confirm the number is correct before pressing the "EXIT" key
to close the output and save the input data. If the data is not saved, press the "Reset" switch to
turn off the output.
9.9 Insulation resistance 0.5M
Ω
correction:
Press the "SET" key, the program enters the insulation resistance correction mode of 0.5M
Ω
,
and the display will show:
Connect a standard resistance of about 500K
Ω
/ 2W to the "H.V." and "terminal under test" on
this instrument, and then press the "Start" key. At this time, the calibration program of this
instrument will automatically output a voltage of about 1000VDC , The display will show:
Please use the
“
+
”
or
“
-
”
key on the panel to input the standard resistance into the calibration
program, the unit is M
Ω
. Please confirm the number is correct before pressing the "EXIT" key
to close the output and save the input data. If the data is not saved, press the "Reset" switch to
turn off the output.
9.10 Insulation resistance 5M
Ω
correction:
Press the "SET" key, the program enters the insulation resistance 5M
Ω
correction mode, and
the display will show:
Connect a standard resistance of about 5M
Ω
/ 0.25W to
“
H.V.
”
and
“
terminal under test
”
on this instrument, and then press the
“
Start
”
key. At this time, the calibration program of this
instrument will automatically output a voltage of about 1000VDC. Voltage, the display will
show:
Please use the
“
+
”
or
“
-
”
key on the panel to input the standard resistance into the calibration
program, the unit is M
Ω
. Please confirm the number is correct before pressing the "EXIT" key
to close the output and save the input data. If the data is not saved, press the "Reset" switch to
CAL IR R1 = 0.500M
Ω
<TEST> to Calibrate
CAL IR R1 = X
.
XXXM
Ω
<EXIT> to Save
CAL IR R2 = 5.000M
Ω
<TEST> to Calibrate
CAL IR R2 = X
.
XXXM
Ω
<EXIT> to Save