29
Please use the "+" or "-" key on the panel to input the standard voltage into the calibration
program, press the "+" key to increase the number, and the "-" key to decrease the number.
Please confirm the number is correct before pressing the "EXIT" key to close the output and
save the input data. If the data is not saved, press the "Reset" switch to turn off the output.
9.3 Correction of AC withstand voltage 10mA
Press the "SET" key, the program enters the AC voltage withstand current 10mA correction
mode, and the display will show:
Please connect the standard AC ammeter and a resistor of about 100K
Ω
/ 10W in series first,
and then connect it between the
“
H.V.
”
and the
“
measured end
”
of the instrument. The
ammeter depends on the
“
measured end
”
. Please press the "Start" key, and the calibration
procedure of this instrument will automatically output a current of about 1000VAC / 10.00mA,
and the display will show:
Please use the "+" or "-" key on the panel to input the standard value of the standard ammeter
into the calibration program, the unit is mA. Please confirm the number is correct before
pressing the "EXIT" key to close the output and save the input data. If the data is not saved,
press the "Reset" switch to turn off the output.
9.4 Correction of AC withstand voltage 2mA
Press the "SET" key, the program enters the AC withstand voltage and current 2mA correction
mode, and the display will show:
Please connect the standard AC ammeter with a resistor of about 500K
Ω
/ 2W in series first,
and then connect it between the
“
H.V.
”
and the
“
measured end
”
of the instrument. The
ammeter rests on the
“
measured end
”
side. Please press the "Start" key, the calibration
procedure of this instrument will automatically output a current of about 1000VAC / 2.000mA,
and the display will show:
CAL ACW I1 = 10.00mA
<TEST> to Calibrate
CAL ACW I1 = XX
.
XXmA
<EXIT> to Save
CAL ACW I2 = 2.000mA
<TEST> to Calibrate
CAL ACW I2 = XX
.
XXmA
<EXIT> to Save