24
After setting the output frequency and pressing the "SET" key, the program will enter the arc
sensitivity setting mode, and the LCD will display:
AC withstand and DC withstand voltage mode
Please use the "+" or "-" key on the face plate to input the arc sensitivity value to be set.The arc
sensitivity is divided into 1-9, there are 9 segments to set, 9 is the highest sensitivity, 1 is the
lowest sensitivity, 0 is the arc condition of the object to be detected. When the range of X is 1-9,
the corresponding arc peaks are: 20mA, 18mA, 16mA, 12mA, 10mA, 7.7mA, 5.5mA, 2.8mA,
and the factory default setting is 0.
Note: This mode is only available for AC withstand voltage and DC withstand voltage.
12. Connect test settings
After the AC / DC withstand voltage arc sensitivity or insulation resistance delay judgment time
is set and the
“
SET
”
key is pressed, the program will enter the connection test setting mode,
and the LCD will display:
Please use the "+" or "-" key on the panel to select the connection test as "YES" or "NO". If the
connection test is set to "YES", after this test is completed, it will automatically connect to the
next group to continue the test. The maximum number of connections is 3 times. If set to "NO",
the test will stop immediately after the completion of this test, and will not be connected to the
next group of tests. When the connection test is set to "YES", the program will automatically
display the symbol "_" after the group, which means that it will connect to the next group of
tests after this group of tests.
This is the last step of parameter setting. You can press the "SET" key again to return to the first
parameter setting step. Press the "EXIT" key to save the data or press the "Reset" key to save the
data without leaving the parameter setting mode. The program automatically calls the test
parameters of the current group, enters the test mode, and prepares for testing.
8.3 LCD information
The following are the various types of information that will appear on the display when the
instrument performs AC / DC withstand voltage or insulation resistance tests, as described
below.
1. Test and parameter setting mode (SETUP)
The following display information indicates that the instrument has entered the test and
parameter setting mode of AC / DC withstand voltage or insulation resistance or ground
resistance:
AC withstand voltage test
DC withstand voltage test
ACW SETUP
XXX.XS
or
DCW SETUP
XXX.XS
CONNECT = YES
Select by
∧ or ∨
Arc Sense= X
Range:0-9
0=OFF