Setting the beeping steps

5.1.3 [Nominal value] input

The absolute and relative values must be entered in a nominal value. Direct reading value

comparison method The nominal value does not participate in the calculation, but in the [standard

[Setup] Setting display

Nominal] In the range mode, the nominal value of the resistance will participate in the range

selection.

Method, you need to enter the correct nominal value of the resistor.

Enter the nominal value:
Step 1

Select the [Measurement Settings] function key in the upper bar to

switch to the <Measurement Settings> page

Step 2

Use the cursor keys to select the [Nominal resistance] or [Nominal

voltage] field;

Step 3

Use the numeric keys to enter data, and the unit to select using the

function keys.

5.1.4 [Lower Limit] and [Upper Limit] settings

Step 1

Select the [Measurement Settings] function key in the upper bar to

switch to the <Measurement Settings> page

Step 2

Use the cursor keys to select the [Lower Resistance] or [Lower

Voltage] field;

Step 3

enter data using the numeric keypad

Step 4

Use the cursor keys to select [Resistance Upper Limit] and

[Resistance Lower Limit]

Step 5

enter data using the numeric keypad

Step 6

Repeat 2 ~ 5 to complete the data input of other files.

Enter data using the numeric keypad, then select the unit to complete the operation.

Note: The instrument shares the same storage space for the three comparison modes, so the

comparator data needs to be reset after switching the comparison mode

6. System settings

Close

Off beep

Pass

The buzzer sounds when the sorting

result is pass (GD)

Failed

The buzzer sounds when the sorting

result is disqualified (NG

Summary of Contents for JK625

Page 1: ...u Jinailian Electronic Technology Co Ltd No C3 Building 22 New Impetus Pioneering Center No 1 Qingyang North Road Tianning District Changzhou Jiangsu CN TEL 0086 519 85563477 FAX 0086 519 85565067 https jinailian en alibaba com www jk17 com www jaldz com JK625 JK625L ...

Page 2: ...ument in explosive atmospheres Do not use the instrument in a flammable or explosive atmosphere steam or dusty environment Using any electronic device in such an environment is a personal risk Do not open the instrument case Non professional maintenance personnel must not open the instrument case in an attempt to repair the instrument There is still a clean charge that has not been discharged for ...

Page 3: ...pinion has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling For warranty service or repair this product must be returned to a service facility designated by Jinailian The buyer shall prepay shipping charges to Jinailian and Jinailian shall pay shipping charges to return the product to the Buyer However the Buyer shall pay all shipping charges du...

Page 4: ...quirements 6 1 3 Operating Environment 6 1 4 Cleaning 6 1 5 Replacing the battery 6 1 6 Adjusting Support 7 2 Overview 8 2 1 Introduction 8 2 2 Main Specifications 8 2 3 Main functions 8 2 3 1 Calibration function 8 2 3 2 Comparator Function Sorting Function 8 2 3 3 System Settings 9 2 3 4 Remote Control 9 3 Start 9 3 1 Front Panel 9 3 2 Interface Panel 10 3 3 Using an external power supply 10 3 3...

Page 5: ...7 5 1 3 Nominal Value Input 17 5 1 4 Setting of Lower Limit and Upper Limit 18 5 2 Clearing the short circuit 18 6 System Configuration 18 6 1 System Configuration page 18 6 1 1 System Language 19 6 1 2 System Account Password 19 6 1 3 System Date Time 20 6 1 4 Sound Settings 20 6 1 5 Baud rate setting 20 6 2 System Information page 22 7 Remote Communication 22 7 1 USB 232C 22 7 1 1 USB 232C conne...

Page 6: ...y use our special AC power adapter and lithium battery AC power adapter Input voltage 90V 260VAC 49Hz 62Hz Power up to 10VA Warning Do not use power adapters of other specifications The instrument can only use our company s power supply and rechargeable lithium battery 1 3 Operating environment JK625 must be used under the following environmental conditions Temperature 0 55 Humidity less than 70 R...

Page 7: ...ew battery paying attention to the direction of the plug 3 Insert the new battery into the battery compartment close the battery cover and tighten the screws 1 6 Adjusting the support The instrument supports 1 position for user convenience 60 degrees Use the 60 degree position to make the instrument more stable Figure 1 2 Support position at 60 degrees ...

Page 8: ...rting results show the sorting sound setting which is applied to the UPS online battery measurement automatic sorting system to complete a fully automatic assembly line test JK625 is equipped with 500 sets of data storage which can save measurement data with one click saving you the time of recording data And can be equipped with a USB interface for remote control and data acquisition and analysis...

Page 9: ... Sound setting Users can set OFF GD NG sound according to their needs 2 3 3 System Settings 1 Serial switch function 2 Switch between Chinese and English 3 Date and time settings 4 I O port switch setting 5 U disk switch setting 6 Automatic shutdown time setting 2 3 4 Remote control Supports a maximum baud rate of 115200bps compatible with SCPI protocol and ASCII transmission 3 Get started In this...

Page 10: ...r adapter cannot be replaced It is recommended to use our company s special power supply Figure 3 2 Connecting the external power adapter to the instrument Plug the power adapter into the AC adapter input jack on the instrument 3 3 1 Battery charging function If the battery is not fully charged the instrument s charging circuit will automatically start to charge the internal lithium battery after ...

Page 11: ...ts We recommend that you use our test fixtures or test cables The contact springs of our test fixtures are silver plated or gold plated which will cause abrasion of the electroplated surface during long term work such as 1 2 years It is recommended that you replace the new fixtures in time if you find that there are obvious deviations in several tests 4 Meas Measurement display In this chapter you...

Page 12: ...er mode steps Step 1 Select the Display function key in the upper column to switch to the Measurement Display page Step 2 Use the cursor keys to select the Trigger field Step 3 Use the Enter function key to select the trigger mode Function key Function Automatic trigger mode Manual trigger mode 4 1 2 Measurement Range The instrument has 3 measuring ranges automatic range manual range and nominal r...

Page 13: ... range Range number Range definition Resistance measurement range 3 3Ω 320mΩ 3 3Ω 2 3mΩ 0mΩ 3 3mΩ 1 30mΩ 32mΩ 330mΩ 0 300mΩ 32mΩ 330mΩ There are 3 measuring method Range method Description Advantages Disadvantages Automatic The instrument automatically selects the best test range according to the impedance value and the range number in the range field is set automatically The user does not need an...

Page 14: ...he automatic range method is not suitable For sorting users please select the nominal range method 4 1 3 Upper and lower limit settings of resistance and voltage 4 1 4 Information bar on Measurement Display page Below the instrument test results the instrument also displays the sorting results and related parameters These information are modified in the Comparer Settings main page and are here for...

Page 15: ... icon there is a change in charge indicating that charging is in progress no change indicates that charging is complete USB flash drive is currently inserted H The current data is held T Temperature compensation is on not available in AT518L 20 0 example Measured temperature not available in AT518L 4 2 Short circuit clear Press the Display shortcut key and then press the Short Circuit Clear functi...

Page 16: ...lay main page Step 2 Press the Short Circuit Clear function key in the upper bar again Cleared in about 3 seconds the screen refreshes and flashes 5 Setup setting display In this chapter you will learn about all the setting functions Measurement Settings page 5 1 Measurement Settings page At any time as long as you press the Measurement Settings shortcut key you can enter the Measurement Settings ...

Page 17: ...unction key in the upper bar to switch to the Measurement Settings page Step 2 Use the cursor keys to select the Speed field Step 3 Select the speed you want to set Function key Function Slow test speed is slow medium speed Test speed is medium speed Fast speed Test speed is fast Top speed Test speed is top speed 5 1 2 Sound The beep function is effective only when the comparator function is turne...

Page 18: ...unit to select using the function keys 5 1 4 Lower Limit and Upper Limit settings Step 1 Select the Measurement Settings function key in the upper bar to switch to the Measurement Settings page Step 2 Use the cursor keys to select the Lower Resistance or Lower Voltage field Step 3 enter data using the numeric keypad Step 4 Use the cursor keys to select Resistance Upper Limit and Resistance Lower L...

Page 19: ...gs on the System Configuration page will be automatically saved in the system and will be loaded automatically at the next boot Figure 6 1 System Configuration page 6 1 1 Serial switch The instrument supports two states of serial port opening and closing Set the serial port switch steps Step 1 Select the System Settings button in the task bar to enter the System Settings page Step 2 Use the cursor...

Page 20: ...top bit 1 bit Parity None Baud rate configurable Set the baud rate steps Step 1 Select the System Settings button in the task bar to enter the System Settings page Step 2 Use the cursor keys to select the Baud Rate field Step 3 Select the baud rate to be set Function key Function 2400 4800 9600 14400 19200 6 1 3 U disk switch The instrument supports two states of U disk open and close Steps of set...

Page 21: ...utton in the task bar to enter the System Settings page Step 2 Use the cursor keys to select the Language field Step 3 Press the enter key to set the language Function key Function CHINESE CHINESE ENGLISH ENGLISH 6 1 6 System Date Time The instrument uses a 24 hour clock Set date steps Step 1 Press the display button to enter the main page Step 2 Select the System button in the task bar to enter t...

Page 22: ...ey Function Hour 1 hour Hour 1 hour Minute 1 minute Minute 1 minute Second 1 second Second 1 second File name 6 2 System Information page The System Information page has no user configurable options Figure 6 2 System Information Page View system information steps Step 1 Select the System Information button in the task bar to enter the System Settings page Step 2 Select the System Information butto...

Page 23: ...ter connecting the instrument please find the USB driver on the CD ROM installation 7 1 1 USB 232C connection The instrument has built in USB 232 communication Before use just plug the USB cable into the instrument and use USB to communicate with the computer Default communication settings of the instrument Transmission method full duplex asynchronous communication with start and stop bits Data bi...

Page 24: ...truments for testing instruments SCPI also known as TMSL Test and Measurement System Language was developed by Agilent Technologies based on the IEEE 488 2 extension and has been widely adopted by test equipment manufacturers to date See The instrument s built in command parser is responsible for the user s various command format parsing Since the command parser is based on the SCPI protocol but i...

Page 25: ... manual range Maximum reading Resistance 3300 Voltage 60 000 Range mode automatic manual and nominal Calibration short circuit full range clear Test terminal Four terminal test method Custom sorting sound Comparator ABS PER and SEQ Handler 8 levels in total 3 levels of resistance HI IN LO 3 levels of voltage HI IN LO 1 level is open 1 level is always out of range Beep off pass fail Trigger Interna...

Reviews: