20/39
*The steps to set up the other channel upper limit and lower limit are the same as above.
Steps to switch channel pages :
First steps
Press [settings] to enter <function Settings > main page
Second steps
Press the [sorting settings] key to enter <sorting Settings > page.
Third steps
Press the function key [Page] to switch the page.
5.3 User correction
Press the [setup] key, and then press the function key [calibration] to enter the < user
correction> page.
Users can complete the correction settings of the channel data under this page.
Figure 5- 3 <User correction>page
5.3.1 [001]
Steps to correct the 001 channel:
First steps
Press [settings] to enter <function Settings > main page
Second steps Press the [calibration] key to enter the <user correction> page.
Third steps
Use cursor keys to select [0.0] fields.
Fourth steps
Using function keys to select
Fifth steps
Function keys
Function
Input correction
value
Input the correction temperature value of the
selected channel, enter data with digital keyboard,