The third step is to use function key selection.
Function key
Function
Closed
Beep closed
Qualified beep
When the sorting result is qualified (GD), the buzzer calls.
Unqualified beep
When the sorting result is unqualified (NG), the buzzer calls.
6.1.5 [baud rate] settings
The instrument has built-in RS-232 interface. After sensing the signal transformation of
RS-232 interface, the instrument immediately communicates with the host computer
according to the set baud rate.
At the same time, the keyboard is locked.
In order to communicate correctly, please make sure that the baud rate is set correctly.
The difference between the upper computer and the instrument will not communicate
correctly.
The instrument RS-232 is programmed in SCPI language.
The RS-232 configuration is as follows:
Data bits: 8 bits
Stop bit: 1 bits.
Parity check: None
Baud rate: configurable
Set baud rate:
The first step is to enter the < system configuration > page.
The second step uses the cursor key to select the baud rate field.
The third step is to use function key selection.
Function key
Function
1200
If you use a communication converter with optocoupler isolation, please
use this baud rate.
9600
38400
57600
115200
It is recommended that you use this high speed baud rate for
communication with the host computer.
6.1.6 Communication [instruction handshake] switch
Instrument supports RS232 instruction handshake.
When the instruction handshake is opened, all instructions sent by the host to the
instrument will be returned to the host as they are, and then the data will be returned.
After the command handshake is closed, the instructions of the host Send out to the
instrument will be processed immediately.
The steps to set the instruction handshake are:
The first step is to enter the < system configuration > page.