![Jinko JK2520B User Manual Download Page 16](http://html.mh-extra.com/html/jinko/jk2520b/jk2520b_user-manual_2018511016.webp)
16
Internal
Internal triggering
manual
manual triggering
external
external triggering
remote
remote triggering
4.1.2[range]
SCPI Command: FUNCtion:RANGe {<range number>,min,max}
SCPI Command: FUNCtion:RANGe:MODE {AUTO,HOLD,NOMinal}
SCPI Query Command: FUNCtion:RANGe?
SCPI Query Command: FUNCtion:RANGe:MODE?
JK2520C has 6 ranges.
The range of variation for each range is as follows:
Table
4
‐
1 Range of change
Range number
Range name 10kΩ
Range 3.2kΩ~33kΩ
6
1kΩ
320Ω
~
3.3kΩ
5
100Ω
32Ω
~
330Ω
4
10Ω
3.2Ω
~
33Ω
3
1Ω
320mΩ
~
3.3Ω
2
100mΩ
32mΩ~ 330mΩ
1
10mΩ
<33mΩ
JK2520B have 4 ranges. The range of variation for each range is as follows:
Table
4
‐
2 Range of change
Range number
Range name
Range
4
10Ω
3.2Ω
~
33Ω
3
1Ω
320mΩ
~
3.3Ω
2
100mΩ
32mΩ~330mΩ
1
10mΩ
<
33mΩ
There are 3 ways of measuring.
Table 4 - 3 test range specification
Range
mode
Description
Advantage
Disadvantage