All the settings related to measurement are operated in < Settings > page.
< Settings > page, the instrument does not display test results or comparator results, but
the instrument test is still in progress.
These settings include the following parameters:
Trigger -trigger mode
Range - Test Range
Speed - test speed
Beep-Beep -alarm
Resistance,voltage Comparison method
Resistance nominal value and voltage nominal value
Upper and lower limit setting
The settings of [trigger], [range], [sound] and [speed] can also be set on the <
measurement display > page. For the setting of these parameters, please refer to the
[Meas] main page section.
Fig 5
‐
1< Settings > page
5.1.1 Comparator settings
SCPI Command: COMP:RMOD {OFF,ABS,PER,SEQ}
SCPI Command: COMP:VMOD {OFF,ABS,PER,SEQ}
The instrument can compare resistance and voltage at the same time or separately.
There are 3 ways to compare: absolute value comparison (delta).
Percentage comparison (delta%)
Sequential comparison (SEQ)
Absolute value = measured value- nominal value
Percentage %=(measured value -nominal value) / nominal value x 100%
Direct-reading SEQ comparisons use direct-reading measurements compared to the
upper and lower limits of the file, so that nominal values are not required to participate in
the operation.
The steps to set the comparator are:
The first step is to go to < Settings > page.
The second step uses cursor keys to select [resistance comparison] or [voltage]
comparison fields.
The third step is to use function key selection.