58
Technical data
10.2 Display
values
therapy
status
range
Parameter
Range
(step)
Accuracy
Refresh rate
RR
Range: 0-255 bpm
Step: 1 bpm
± 1 bpm
Each breath
I:E
Range: 1:0.0 to 1:9.9
Step: 0.01
± 0.01
Each breath
Vi
Range: 0.0-65.535 l
Step: 0.001 l
± 150 ml
Start of expiration
MV
Range: 0-65.535 l
Step: 0.1 l
± 150 ml (+ mean
value formation)
Start of expiration
LK
Range: 0-250 l/min
Step: 1 l/min
± 3 l/min (+ mean
value formation)
Marginal condi-
tion, dependent
upon the leakage
and leakage
change
Pressure
(hPa/cm H
²
O)
Range: -5-40 hPa/cm
H
²
O
Step: 0.1 hPa/cm H
²
O
IPAP: ± 0.4 hPa
EPAP: ± 0.4 hPa
CPAP: ± 0.4 hPa
2 ms (+ mean
average forma-
tion)
Flow
Range: -247-247 l/
min
Step: 1 l/min
± 1.1 l/min or ±
20%
5 ms