Using the device daily
19
4.4.1 Pausing
alarms
• You can switch the acoustic alarm to pause for the duration of two minutes
by pressing the "Audio pause" key.
• After a period of two minutes the device reactivates the acoustic signal if
the triggering event is still present.
• When you press the "Audio pause" key again during the activated 2-min-
ute counter the duration is reset to two minutes, provided there is still a
triggering event. The counter is deactivated if the triggering event has
been rectified in the meantime.
• The "Audio pause" key may also be pressed as a matter of precaution if
there is a foreseeable condition for an alarm, e.g. before secretion is taken
from a patient. At this point the alarms are switched to pause for two min-
utes.
You can check a more accurate list of the alarms and error messages on
Page 38.