56
Technical data
Performance features:
Max. working pressure:
Min. working pressure:
Max. stable limit pressure:
Min. stable limit pressure:
30 hPa (by pressure measurement/
adjustment)
3 hPa (by pressure measurement/
adjustment)
40 hPa
0 hPa
Maximum respiratory resistance in the
event of single failure:
Inspirational pressure on the patient
connection opening of the device for a
flow of 60 l/min:
Expiration pressure on the device's
patient connection opening for a flow of
60 l/min:
4.05 hPa
4.18 hPa
Inspiration trigger:
Flow-based inspiration trigger
Accuracy: ± 20%
Expiration trigger:
The expiration trigger is triggered dur-
ing the following percentages of the
maximum flow during inspiration:
30 - 70% of the IPAP value
Accuracy: ± 20%
Pressure measurement:
Integrated pressure converter in the
respiratory device
Accuracy: ± 0.4 hPa
Short-term pressure stability
in accordance with DIN EN ISO 17510-
1:2009:
at 3 hPa: ± 0.5 hPa
at 10 hPa: ± 0.5 hPa
at 17 hPa: ± 0.5 hPa
at 23 hPa: ± 1 hPa
at 30 hPa: ± 1 hPa
Long-term pressure stability
in accordance with DIN EN ISO 17510-
1:2009:
< 0.1 hPa