after finishing high temperature setting. Please press “Temperature
Setting Key” when “LO” is flashing, and then adjust the low
temperature setting. The temperature standard is 0.5
℃
a time such
as 34
℃
to 34.5
℃
then 35
℃
then 35.5
℃
and so on and then press
“Temperature setting key” to complete the setting.
Remark:
‧
The range of setting: The default figures of the temperature alarm setting
are: Hi: 38
℃
(104
℉
) and Lo:34
℃
(96
℉
). The high temperature alarm
range from 36
℃
to 40
℃
(96
℉
-104
℉
). The low temperature alarm range
from 32
℃
-35.5
℃
(89
℉
-96
℉
).
‧
Faulty-prevention mechanism: If the above steps setting are not
completed, the Receiver will return to the temperature monitoring mode
without changing any setting within 10 seconds.
‧
Priority alarm processing: The product is designed to process the alarm
situation prior to the High/Low temperature alarm setting mode, or any
other operations. Without saving any setting if not pressing “Temperature
setting key”, the user is better off to operate the temperature alarm setting
with the full power supply.
‧
Please re-set the Hi/Lo temperature alarm figures after replacing the
batteries. The Receiver will take the default as the temperature alarm
setting after battery replacement.