JETT BTM-D1D User Manual Download Page 15

Q: Cannot turn on the Receiver?
A:

1.  Please read carefully the manual again.
2. Check the battery well to make sure batteries are in it, or please 

change new batteries.

3.  Batteries  may be not installed correctly.
4. This product may be out of order, please call to customer center.

Q: The armband power indicator flashing with 

green and red light?

A:

1. Press the power key of armband, the red light will light on; and then 

press the power key again, the green light will turn on. And then press 
the power key for 2secs until red light is on to power off. Now re-start 
the Armband. (Press the power key for 2 seconds until green light is 
flashing.)

2. Replace the battery and start the device again.

Summary of Contents for BTM-D1D

Page 1: ...Body Temperature Monitor User Manual 02 Introduction of BTM D1D Wireless Body Temperature Monitor 03 Operating Instructions 06 The Instructions of Indicators 11 Importance Precautions 13 Q A 14 Technical Specifications 16 Table of Content ...

Page 2: ...2032 X 1 Arm belt X 1 Assisted arm belt X 1 Quick Guide Card X 1 Manual X 1 Warranty CardX 1 General Body Temperature Reference Ear Temperature 35 8 38 0 ArmpitTemperature 34 7 37 3 Oral Temperature 35 5 37 5 Anus Temperature 36 6 38 0 BTM D1DWireless Body Temperature Monitor User Manual ...

Page 3: ...ing key To increase the Hi temperature alarm setting and to switch the Lo setting key To decrease the Lo temperature alarm setting Power indicator To indicate the status of power Beeper LCD Screen LO Setting Key HI Setting Key Switch Temperature Setting Power Key Power Indicator AAA batteres X 2 Introduction of BTM D1D Wireless Body Temperature Monitor ...

Page 4: ... Armpit Temperature Mode Receiver Battery Indicator Armband Battery Indicator Power Key Power Indicator Temperature Sensor CR2032 Battery Front side Back side Battery Installation 1 Tur n the c ounter clockwise by using a coin or a screw driver to open the battery cover 2 Insert the cell battery 3 Turn the clockwise to close the battery cover ...

Page 5: ... poor quality or years old CR2032 battery it may easy to become powerless when power on which means the battery can not supply power normally How to assemble the armband How to wear the armband 1 Wear the armband on the left arm 2 Adjust the belt at the proper tightness 3 Move the temperature sensor under the armpit 4 Wear the assisted arm belt for fixing the sensor under the armpit to attach 1 2 ...

Page 6: ... Step 4 Read the temperature on the LCD screen to check the change of the patient s temperature It takes 3 5 minutes to have stable temperature monitoring of armpit The temperature sensor will update armpit temperature every 12 seconds to be displayed on LCD screen POWER OFF F Step 1 To turn the Receiver off press and hold the Power key for 2 seconds until it beep to turn off the LCD screen of the...

Page 7: ...armpit temperature status When the temperature is out of the setting range the Receiver will start a series of Beeps alerting the caregiver to take appropriate actions The alarm setting procedures must be done when the armband is on Follow the steps below Step1 High temperature setting Press the Temperature setting key for 2 seconds the Receiver will beep twice to get into the temperature alarm se...

Page 8: ...rom 32 35 5 89 96 Faulty prevention mechanism If the above steps setting are not completed the Receiver will return to the temperature monitoring mode without changing any setting within 10 seconds Priority alarm processing The product is designed to process the alarm situation prior to the High Low temperature alarm setting mode or any other operations Without saving any setting if not pressing T...

Page 9: ...ading of the patient is out of the temperature alarm setting range That is the Hi and its figures are flashing The Receiver will beep every 2 seconds to inform the caregiver to take appropriate action Anti sleepyhead mechanism The Receiver will check every 10 minutes the temperature of the patient if it is still in the temperature alarm setting The Receiver will The Hi Lo setting ...

Page 10: ...every 10 minutes the caregiver can set the Hi Lo temperature alarm a little bit higher lower and then result the setting after the patient is getting better In case that the temperature when power on lowers the low temperature alarm setting then starts the alarm system this product will trigger the low temperature alarm system after normal temperature higher than the low temperature setting by 0 5...

Page 11: ...power off Orange Low Volt Every 4 seconds when power on Red No Signal Every 10 seconds when power on Power Supply Status LCD Screen Battery Display Alarm Siren Full power supply Full battery on Armband Battery Indicator No siren Low power supply Flashing on Armband Battery Indicator Beeps twice every 10 seconds Power Supply Status LCD Screen Battery Display Alarm Siren Full power supply Full batte...

Page 12: ...g for the screen in monitoring mode Otherwise the armband will automatically switch off after 3mins If the problem can not be solved from the above suggestion please restart both Armband and Receiver again 5 Error Solution LCD Screen Meaning Solution E r 0 Connection Abnormal System breakdown needed repair Status Temperature Indicator Alarm Siren Poor signal Beeps twice every10 seconds 12 Beep Bee...

Page 13: ...e abnormal temperature range The measurement result however could sometimes be influenced by the surrounding temperature The measurement result is ONLY for reference Please seek qualified physician s help if necessary 8 Please pay much attention on children with using this product Please also keep it out of reach of children and place it on a higher position 9 Please do not use any rechargeable ba...

Page 14: ...wer supply please change batteries 4 Make sure you are not using this product directly after exercising bathing etc please put the product for 30 minutes at room temperature and then wipe it up 5 Possible interference please turn the product power on again 6 Out of order contact customer service immediately Q If allergy is evident after using this product A Please stop using it Q Failure to use th...

Page 15: ...er please call to customer center Q The armband power indicator flashing with green and red light A 1 Press the power key of armband the red light will light on and then press the power key again the green light will turn on And then press the power key for 2secs until red light is on to power off Now re start the Armband Press the power key for 2 seconds until green light is flashing 2 Replace th...

Page 16: ...ility regulation for medical device EN12470 4 EN12470 5 ISO 13485 2003 ISO 9001 2000 ISO 10993 1 2003 Receiver Armband Power supply Two AAA batteries Up to 80 100 hours of constant operation One 3Volt lithium cell battery CR2032 Up to 72 96 hours of constant operation If under airwave interference environment the battery life will be shortened Dimensions mm 130H X 50W X 32D 136H X 85W X 9 5D Weigh...

Reviews: