2. If outside of signal area and then result in disconnection, please remove
the receiver close to the armband within 3 min and then the signal will be
connected.
3. Any abnormal occurred when turn on the armband, please check and
change the battery.
4. Recommend turning on both Receiver and Armband power within 20
seconds for connection, otherwise it will automatically power off.
(2)How to set the temperature alarm:
Before using the body temperature monitor function, the operator may need
to set both high and low temperature alarm figures according to the patient’s
general armpit temperature status. When the temperature is out of the setting
range, the Receiver will start a series of "Beeps" alerting the caregiver to take
appropriate actions. The alarm setting procedures must be done when the
armband is on. Follow the steps below:
Step1: High temperature setting: Press the “Temperature setting key” for 2
seconds, the Receiver will beep twice to get into the temperature
alarm setting mode. The LCD screen flashes “HI” and the original
figures will be displayed.
Step2: Press “Temperature Setting Key” to adjust the alarm temperature
setting. The temperature standard is 0.5
℃
a time such as 38.5
℃
to
39
℃
then to 39.5
℃
and so on, then press “Temperature setting key”
to complete the setting.
Step3: Low temperature setting: low temperature setting starts automatically