8
SECTION 1 - Introduction
1.1
INSTRUMENT DESCRIPTION
The 7310 and 7315 spectrophotometers are suited to a wide range of applications in education,
quality control, environmental and clinical analysis. The 7310 is a visible spectrophotometer covering
a wavelength range from 320nm to 1000nm. The 7315 is a UV/Visible spectrophotometer with a
wavelength range from 198nm to 1000nm. Both models have five measurement modes: photometrics,
concentration, spectrum scanning, quantitation and kinetics. These instruments use icon driven software
and have an improved navigation system for easy and intuitive usability.
1.2
INSTRUMENT SPECIFICATION
7310
7315
Wavelength
Range
320 to 1000nm
198 to 1000nm
Resolution
1nm
Accuracy
± 2nm
Repeatability
± 0.5nm
Spectral bandwidth
5nm
Photometrics
Transmittance
0 to 199.9%
Absorbance
-0.300 to 2.500A
Accuracy
±1%T, ±0.01Abs at 1.000 Absorbance
Resolution
0.1%T, 0.001A
Stray light <0.5% at 340nm <0.5% at 340nm and 220nm
Stability
<0.002Abs/hr after 30 minute warm up
<0.001Abs/hr without warm up
Concentration
Range
-300 to 9999
Resolution
Selectable 1/0.1/0.01/0.001
Calibration
Blank with a single standard or factor
Units no units, %, ppm, EBC, SRM, mEq/l, mEq, M, mM, µM, nM, U, U/l, U/ml, g/l, mg/l,
µg/l, ng/l, g/dl, mg/dl, µg/dl, mg/ml, µg/ml, ng/ml, µg/µl, ng/µl, mol/l, mmol/l
Factor
0.001 to 10000
Standard
0.001 to 1000
Quantitation
Range
-300 to 9999
Resolution
Selectable 1/0.1/0.01/0.001
Calibration
Blank with up to 6 standards
Units no units, %, ppm, EBC, SRM, mEq/l, mEq, M, mM, µM, nM, U, U/l, U/ml, g/l,
mg/l, µg/l, ng/l, g/dl, mg/dl, µg/dl, mg/ml, µg/ml, ng/ml, µg/µl, ng/µl, mol/l, mmol/l
Curve fit algorithms Quadratic, quadratic through zero, linear, linear through zero, interpolate
Kinetics
Measurement Time
2 to 9999 seconds
Calibration
Blank with a single standard or factor
Display Concentration, rate of change, initial and final absorbance/%T
Resolution
Selectable 1/0.1/0.01/0.001