undertaken by personnel trained and equipped to verify the
optical performance of the instrument.
Instrument covers should only be removed by engineers
trained in safe working practices and aware of electric shock
hazards. Removal of the monochromater cover will invalidate
any warranty claim regarding performance to specification.
Do not look directly at the light source, use eye protection or
the lamp dimming function when necessary. Do not touch any
optical surfaces.
If in any doubt DO NOT PROCEED.
Ensure or verify by replacement that a GENUINE JENWAY
LAMP is fitted, (refer to the instruction manual for details).
There are many lamps that look similar but the filament
position is critical and can only be guaranteed on Jenway
lamps, replacing the lamp may correct the problem without
any further re-alignment.
1
Remove the top half of the case by undoing the four recessed
screws in the corners of the base.
2
Place the top half of the case behind the base taking care not to trap
or strain the connecting cables.
3
Remove the two screws retaining the black monochromator cover
in the bottom half of the case. Remove the cover and place to one
side
. DO NOT TO TOUCH ANY OPTICAL COMPONENTS.
4
Taking all precautions to avoid the risk of electric shock, connect
the power cable and depress the right arrow key [>] while turning
the power switch on.
5
The display should show a mV reading on the top row, followed by
the wavelength on the 2
nd
row and %T at the bottom left hand side.
6
Pressing the right arrow key again [>] should operate the solenoid
in front of the exit slit assembly. Press this key to ensure that the
shutter is in the open position (light is passing through the exit slit)