7
REPLACE ELECTRODE
1.
Unscrew the electrode collar to remove the electrode &
ATC assembly as shown in the right figure.
2.
Remove the old electrode from
the electrode collar.
3.
Insert a new electrode, make
sure the electrode fit back into
the meter correctly.
4.
Screw back the electrode collar.
5.
Soak the electrode in a pH 4 buffer solution for 10
minutes and recalibrate the pH630FA following the
instructions detailed in section
USING VISIONPLUS
pH630FA..