JENCO VisionPlus pH630FA Operation Manual Download Page 7

6

reading is reached, the “HOLD” icon will be on. Press the

“Enter”

key to save the calibration, wait a second, the

second point has been calibrated and the unit is ready to

be sloped at the third buffer.

[

Note:

At this moment

,

Press the

“Hold”

key, the unit

will exit the calibration mode.

Dual points

calibration is

complete.]

5. Remove the electrode & ATC assembly from the second

buffer. Rinse them in distilled water and immerse them in

the third buffer solution. The unit will display the

temperature of the third buffer.

6. The first, second buffer icons will be on, the third buffer

icons will flash and the “HOLD” icon will flash until the

unit detects a stable reading, When a stable reading is

reached, the “HOLD” icon will be on. Press the

“Enter”

key to save the calibration, wait a second, the third point

has been calibrated and the unit will automatically exit

the calibration mode.

Three points

calibration is

complete.
[

Note:

For accurate measurements, it is recommended

that pH calibration is preformed once a week and after
replacing the electrode.]

D. Measure

Dip the meter into the test solution in the “Measure

Mode”.

E. Hold Data

1.

When the pH reading is stable, press

“Hold”

key once

to lock the reading.

2.

Press

“Hold”

key again to unlock reading and the unit

will return to “Measure Mode”. The unit is now ready for

another measurement.

Summary of Contents for VisionPlus pH630FA

Page 1: ...Operation Manual MODEL pH630FA Microcomputer Based pH Temperature Pocket Meter pH630FA...

Page 2: ...and Meter 2 VISIONPLUS pH630FA OVERVIEW 3 A Meter Description 3 B LCD Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH630FA 5 A Power on...

Page 3: ...can operate within specifications for approximately 2 to 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and inse...

Page 4: ...ATC assembly Electrode cap is not shown B LCD Display 1 Calibration buffer s and number of calibration point 2 Low battery indicator 3 Calibration mode indicator 4 HOLD mode indicator 5 pH reading 6...

Page 5: ...either be 7 00 7 00 4 01 10 01 pH or 6 86 6 86 4 00 9 18 pH B Keypad Operations Key Operation Mode Duration Function Hold Measure 0 second Holds current measurement reading Press again to resume measu...

Page 6: ...Rinse the electrode ATC assembly in distilled water and immerse them in the first buffer solution The temperature displayed is the buffer temperature 2 Press Cal key to initiate calibration the buffer...

Page 7: ...hird buffer icons will flash and the HOLD icon will flash until the unit detects a stable reading When a stable reading is reached the HOLD icon will be on Press the Enter key to save the calibration...

Page 8: ...e the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar 5 Soak the electrode in a pH 4 buffer...

Page 9: ...or 14 00 pH ovr Over Range or Recalibrate ovr 32 0 to 140 F pH CAL a pH OFFSET b pH SLOPE a Offset 7 00 pH mV 100 mV Offset Offset 6 86 pH mV 108 3 mV b New slope ideal slope by 30 Use a new buffer so...

Page 10: ...n pH Buffer Temperature range pH Electrode Offset recognition AUTO 14 2 F to 212 F 32 0 F to 140 F 100 mV at 7 00 pH 108 3 mV 91 7 mV at 6 86 pH pH Electrode Slope recognition 30 at 4 00 4 01 9 18 and...

Page 11: ...e repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and ser...

Reviews: