JENCO VisionPlus pH630FA Operation Manual Download Page 3

2

INITIAL INSPECTION AND ASSEMBLY

Carefully unpack the instrument and accessories. Inspect for

damages made in shipment. If any damage is found, notify your

Jenco

representative immediately. All packing materials should

be saved until satisfactory operation is confirmed.

BEFORE YOUR FIRST USE

A. Replacing the Batteries

Replace the battery when the blinking
low battery indicator “

” appears

on the upper left corner of the LCD

screen. The instrument can operate

within specifications for approximately

2 to 3 hours after low battery indicator

appears.

1.

Shown in the right figure.

2.

Take off the battery cover.

3.

Remove all of the old

batteries and insert a new set

of batteries ensuring the

polarities are correct.

B. Soak the Electrode

1.

Remove the electrode cap covering the VisionPlus

pH630FA meter.

2.

Soak the electrode in a 4 pH buffer solution for 10

minutes before first use or after storage.

C. Setup and Calibrate the Electrode and Meter

VisionPlus pH630FA must be setup and calibrated before your

first use. Please follow the instructions detailed in section

USING VISIONPLUS pH630FA 6

.

1

2

Summary of Contents for VisionPlus pH630FA

Page 1: ...Operation Manual MODEL pH630FA Microcomputer Based pH Temperature Pocket Meter pH630FA...

Page 2: ...and Meter 2 VISIONPLUS pH630FA OVERVIEW 3 A Meter Description 3 B LCD Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH630FA 5 A Power on...

Page 3: ...can operate within specifications for approximately 2 to 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and inse...

Page 4: ...ATC assembly Electrode cap is not shown B LCD Display 1 Calibration buffer s and number of calibration point 2 Low battery indicator 3 Calibration mode indicator 4 HOLD mode indicator 5 pH reading 6...

Page 5: ...either be 7 00 7 00 4 01 10 01 pH or 6 86 6 86 4 00 9 18 pH B Keypad Operations Key Operation Mode Duration Function Hold Measure 0 second Holds current measurement reading Press again to resume measu...

Page 6: ...Rinse the electrode ATC assembly in distilled water and immerse them in the first buffer solution The temperature displayed is the buffer temperature 2 Press Cal key to initiate calibration the buffer...

Page 7: ...hird buffer icons will flash and the HOLD icon will flash until the unit detects a stable reading When a stable reading is reached the HOLD icon will be on Press the Enter key to save the calibration...

Page 8: ...e the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar 5 Soak the electrode in a pH 4 buffer...

Page 9: ...or 14 00 pH ovr Over Range or Recalibrate ovr 32 0 to 140 F pH CAL a pH OFFSET b pH SLOPE a Offset 7 00 pH mV 100 mV Offset Offset 6 86 pH mV 108 3 mV b New slope ideal slope by 30 Use a new buffer so...

Page 10: ...n pH Buffer Temperature range pH Electrode Offset recognition AUTO 14 2 F to 212 F 32 0 F to 140 F 100 mV at 7 00 pH 108 3 mV 91 7 mV at 6 86 pH pH Electrode Slope recognition 30 at 4 00 4 01 9 18 and...

Page 11: ...e repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and ser...

Reviews: