JENCO VisionPlus pH630FA Operation Manual Download Page 5

4

OPERATION MODES AND KEYPAD OPERATIONS

A. Operation Modes

VisionPlus pH630FA meter has 4 operation modes:

1.

Measure Mode. Measure Mode is used to make all pH

and temperature measurements.

2.

Calibration Mode. Calibration Mode is used to perform

1, 2 or 3 point calibration.

3.

Hold Mode. Hold Mode is used to display held measured

values for increased ease of use.

4.

Buffer Select Mode. Buffer Select Mode is used to select

the buffer set, which can either be 7.00(7.00/4.01/10.01

pH) or 6.86(6.86/4.00/9.18 pH).

B. Keypad Operations

Key

Operation

Mode

Duration Function

Hold

Measure

0 second

Holds current

measurement reading.

Press again to resume

measuring.

Hold

0 second Returns to Measure Mode.

Calibration

0 second Leaves Calibration Mode.

E

nter

M

ode

C

al

Measure

0 second Enters Calibration Mode.

Calibration

0 second

The “HOLD” icon is displayed,

then the unit will recalibrate the

buffer.

Measure

5 seconds

or more

Enters Buffer Select Mode.

M

ea.

E

ff

.

S

tand

S

lope

M

ode

C

lear

All

0 second Turns meter on/off.

E

nter

M

ode

C

al

Calibration

0 second

The “HOLD” icon is displayed,

then the unit will save the

calibration.

Summary of Contents for VisionPlus pH630FA

Page 1: ...Operation Manual MODEL pH630FA Microcomputer Based pH Temperature Pocket Meter pH630FA...

Page 2: ...and Meter 2 VISIONPLUS pH630FA OVERVIEW 3 A Meter Description 3 B LCD Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH630FA 5 A Power on...

Page 3: ...can operate within specifications for approximately 2 to 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and inse...

Page 4: ...ATC assembly Electrode cap is not shown B LCD Display 1 Calibration buffer s and number of calibration point 2 Low battery indicator 3 Calibration mode indicator 4 HOLD mode indicator 5 pH reading 6...

Page 5: ...either be 7 00 7 00 4 01 10 01 pH or 6 86 6 86 4 00 9 18 pH B Keypad Operations Key Operation Mode Duration Function Hold Measure 0 second Holds current measurement reading Press again to resume measu...

Page 6: ...Rinse the electrode ATC assembly in distilled water and immerse them in the first buffer solution The temperature displayed is the buffer temperature 2 Press Cal key to initiate calibration the buffer...

Page 7: ...hird buffer icons will flash and the HOLD icon will flash until the unit detects a stable reading When a stable reading is reached the HOLD icon will be on Press the Enter key to save the calibration...

Page 8: ...e the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar 5 Soak the electrode in a pH 4 buffer...

Page 9: ...or 14 00 pH ovr Over Range or Recalibrate ovr 32 0 to 140 F pH CAL a pH OFFSET b pH SLOPE a Offset 7 00 pH mV 100 mV Offset Offset 6 86 pH mV 108 3 mV b New slope ideal slope by 30 Use a new buffer so...

Page 10: ...n pH Buffer Temperature range pH Electrode Offset recognition AUTO 14 2 F to 212 F 32 0 F to 140 F 100 mV at 7 00 pH 108 3 mV 91 7 mV at 6 86 pH pH Electrode Slope recognition 30 at 4 00 4 01 9 18 and...

Page 11: ...e repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and ser...

Reviews: