22
This screen will let the user select what temperature probe (10K
Ω
thermistor, Balco 3K
Ω resistor or RTD pt-1000)
the user will use. This
screen will always default to the last temperature probe the user
selected. For most application the user will only use one kind of
temperature probe so the user don’t need to change this selection
once you select the correct temperature probe. If the user needs to
change the temperature probe type, press the [
¿
] or [
À
] keys to
choose your temperature probe then press [
OK
] to save your new
selection.
ii. OFFSET SETTING screen
This screen selects the OFFSET for
relative ORP. The value here will be
subtracted from the absolute ORP
reading to display the relative ORP
reading. The equation is :
“REL mV” = “ABS mV” – “Offset”
iii. 4mA SETTING screen
This screen will let the user change
the 4 mA value. This value is used
for scaling the 4-20mA
ANALOG
OUTPUT.
(See chapter
VI
.
4-20 mA OUTPUT
.)
iv. 20mA SETTING screen