19
g.
CURRENT SETTING
1.
TRANSMITTER OUTPUT
–
This option would let you choose
if the current output type is
pH
LINEAR, pH ANTILOG or ORP
LINEAR
. If the user chooses
“ORP LINEAR”, the ORP analog
output will be based on the
RELAY/mA UNIT.
2.
4mA OUT
– This value will be used in conjunction with 20 mA to
plot the current output. (See chapter
VI
.
4-20 mA OUTPUT
.)
3
. 20mA OUT
– This value will be used in conjunction with the 4 mA
value to plot the output. (See chapter
VI
.
4-20 mA OUTPUT
.)
Example: For ORP LINEAR: If
4mA OUT
is set at 200AmV,
20mA
OUT
is set at 1000AmV, the analog out at 400AmV=
ANALOG OUTPUT = 4mA + (16mA)/(
1000AmV
-
200AmV
)
=6mA
h. TEMP. CONTROL SETTINGS
1.
RELAY 5
–
The temperature has only
one relay to control you need
to set what control type it will
use,
HIGH
or
LOW
type. (In
HIGH-type, RELAY5 will turn
ON
if the temperature is
greater than the
RELAY5
setting, in LOW-type, the
RELAY5 will turn
ON
if the
temperature is less than the
RELAY5
setting, which is modified by the hysteresis value and
hysteresis mode.
TEMP.CONTROL SETTINGS
1.RELAY 5 :HIGH
2.SET POINT :100.0 C
3.HYSTERESIS:EDGE
* SAVING *
4.HYSTERESIS: 1.0C