14
b.
USER SETTING
You will only see this page if the unit is
not password locked. This page is just
a
WARNING
, informing that all relays
and analog output are frozen, and that
you can calibrate and change the
settings.
NOTE: FROZEN
MEANS ALL THE RELAYS AND THE ANALOG
OUT WILL MAINTAIN THEIR LAST STATE UNTIL THE USER
RETURNS TO MAIN DISPLAY MODE.
c.
pH CALIBRATION
ATC
TEMP
.
–
the
current
temperature of the solution.
1.
BUFFER 1
- in this option you
can select which buffer to use for
the
standardization
calibration.
You can choose
7.00
pH or
6.86
pH by using the
[
]
and
[
/
pH-Recal
.] keys and pressing
the
[
] key to save your choice.
2.
BUFFER 2
- after you selected the buffer 1 this option will let you
select the second buffer to use to calibrate the slope. You can
choose 4.00 pH, 4.01 pH, 9.18 pH or 10.01 pH by using
[
]
and
[
/
pH-Recal
.] keys and pressing the [
] to save your choice.
3.
STAND
- this is the actual pH calibration process, this line will
display the buffer to be used for STAND (OFFSET) calibration ,
depending on the choice you made on
BUFFER 1
. (See Step by
step pH Calibration.)