6 Troubleshooting
6.1 Stop during the operating
Display
Message
Description
Solution
Temperature
Limit
Analog over temperature regulator is
active.
Check the set on the analog over
temperature regulator and the operating.
Restart after temperature is inside of the
range.
Load
Unbalance
Vibration is generating and the
normal shaking is not available. To
prevent the shaking system, stop the
shaking.
Check the unbalanced sample setting,
overload, un-flat ground.
Door Open
Door is opened over 5 minutes
during the operating.
Close the door and touch the start button
and restart the unit.
Sensor Fault
Temperature sensor is faulty so the
operating is stopped.
Contact the manufacturer or sales point.
Over Current
Over current is generated so the
operating is stopped.
Check the overload, obstacle in the shaking
system.